NI PXI-5404 Calibration Procedure
|
© National Instruments
|
5
Although NI-FGEN includes calibration support, the calibration application programming
interface (API) is separate from NI-FGEN. During verification and adjustment, you use
two APIs: the NI-FGEN API and the NI HSSources API. The NI-FGEN API controls the
module both during normal operation and during the verification procedure. The NI HSSources
API controls the module during the adjustment procedure.
NI-FGEN automatically installs the files listed in Table 2.
Table 2.
Calibration File Locations
File Name and Location
Description
<system directory>\niArbCal.dll
The NI HSSources calibration
toolkit library. This toolkit provides
the functionality for calibrating
your module.
VXIpnp\WinNT(Win95)\lib\msc\
niArbCal.lib
VXIpnp\WinNT(Win95)\lib\bc\
niArbCal.lib
A
.lib
file that allows you
to create applications that call
functions in the
niArbCal.dll
:
• If you build an application using
Microsoft Visual C/C++, link to
msc\niArbCal.lib
.
• If you build an application using
Borland C/C++, link to
bc\niArbCal.lib
.
• If you build an application using
LabWindows/CVI, link to the
library appropriate to your
current compatibility mode
(
msc
for Microsoft Visual C/
C++,
bc
for Borland)
VXIpnp\WinNT(Win95)\include\
niArbCal.h
A header file for the accessible
functions in the
niArbCal.dll
.
You must include this file in any
C code that you write to call these
functions.
VXIpnp\WinNT(Win95)\nifgen\
Examples\Calibration\
ExternalCalibrationExample5404.c
Example written in C that illustrates
a calibration procedure using the
functions in the
niArbCal.dll
.