NI PXI-5404 Calibration Procedure
|
© National Instruments
|
43
niHSSources_ApplySessionConstants
Function Prototype
__declspec(dllexport) long __stdcall
niHSSources_ApplySessionConstants(unsigned long
sessionHandle);
Purpose
During calibration, the session holds an internal set of calibration constants. These constants
may have been updated based on calibration steps performed during the session. This function
commits the session constants to the underlying driver so that the module is calibrated according
to the newly updated constants for the remainder of the session. This function does not store the
updated constants in the onboard EEPROM. Calling the function
niHSSources_CalEnd
stores the updated constants in the onboard EEPROM.
Parameters
Name
Description
sessionHandle
A handle to the calibration session for the module; created by calling
niHSSources_CalStart
.