NI PXI-5404 Calibration Procedure
|
© National Instruments
|
45
niHSSources_CalibrateAmplitude
Function Prototype
__declspec(dllexport) long __stdcall
niHSSources_CalibrateAmplitude(unsigned long
sessionHandle);
Purpose
Returns the sine amplitude with the current test value. Use this function during calibration to
iteratively converge on the best value. A new value is calculated, then output is measured, then
a new value is calculated, and so on.
Parameters
Name
Description
sessionHandle
A handle to the calibration session for the module; created by calling
niHSSources_CalStart
.