NI 5122/5124/5142 Calibration Procedure
|
© National Instruments
|
17
16. Move the calibrator test head to the digitizer input channel 1 and repeat steps 2 through 15
for every configuration in Table 3, changing the value of the
channelList
parameter from
"0"
to
"1"
.
17. End the session using the niScope Close VI.
You have finished verifying the vertical gain accuracy of the NI 5122/5124/5142.
Programmable Vertical Offset Accuracy
Complete the following steps to verify the programmable vertical offset accuracy for each
digitizer channel.
1.
Open a session and obtain a session handle using the niScope Initialize VI.
2.
Configure the input impedance and the maximum input frequency using the
niScope Configure Chan Characteristics VI.
10
4.5
-4.5
±0.3975%
±0.65%
20
9
-9
±0.3975%
±0.65%
LabVIEW VI
C/C++ Function Call
Call
niScope_close
with the
following parameter:
vi
: The instrument handle from
niScope_init
LabVIEW VI
C/C++ Function Call
Call
niScope_init
with the
following parameters:
resourceName
: The device name
assigned by MAX
idQuery
:
VI_FALSE
resetDevice
:
VI_TRUE
Table 8.
NI 5124 Vertical Gain Stimuli, Calibration Test Limits,
and Published Specifications
Range
(V)
Positive
Input (V)
Negative
Input (V)
Calibration
Test Limits
Published
Specifications