  NI DCM-23XX Getting Started Guide

What to Do

Complete the following steps to connect power to the DCM.

Ensure the DCM Starter Harness is connected as discussed in the 

Connecting the DCM 

Starter Harness

 section.

Connect power supply leads terminated with properly rated ring terminals to the PDU 
Positive (+) and Negative (-) input studs.

Tighten terminal fasteners to 10.2 N · m (90 lb · in.) using a 1/2-in. wrench.

Connect power supply leads to the power supply. If connecting to a battery, always connect 
the Negative (-) lead last. When disconnecting the battery, always disconnect the 
Negative (-) lead first.

Connecting the DCM to the Host Computer

Complete the following steps to connect the DCM to the host computer or Ethernet network 
using the RJ-45 Gigabit Ethernet port.

Power on the host computer or Ethernet hub.

Connect the RJ-45 Gigabit Ethernet port on the DCM to the host computer or Ethernet hub 
using a standard Category 5 (CAT-5) or better shielded, twisted-pair Ethernet cable.

Caution 

To prevent data loss and to maintain the integrity of your Ethernet 

installation, do not use a cable longer than 100 m (328 ft).

Note 

The DCM attempts to initiate a DHCP network connection the first time you 

connect using Ethernet. The DCM connects to the network with a link-local IP 
address with the form 169.254.x.x if it is unable to initiate a DHCP connection.

Powering on the DCM

When you power on the DCM for the first time, the device boots DCM DSI. The POWER LED 
illuminates, the STATUS LED illuminates briefly, and then the USER1 LED blinks steadily 
once the DSI executable has booted. Overall boot time is approximately 60 seconds.

Configuring the DCM in Measurement and 
Automation Explorer (MAX)

Complete the following steps to find your DCM in MAX.

Ensure that any anti-virus or firewall software running on the host computer allows 
connections to the host computer.

Note 

MAX uses UDP 44525. Refer to the documentation of your firewall software 

for information about configuring the firewall to allow communication through 
UDP 44525.

Launch MAX on the host computer.

Summary of Contents for NI DCM-23 Series

Page 1: ...rds Caution Do not mix hazardous voltage circuits and human accessible circuits in the same cable unless the cables are properly insulated for min 250V Caution Ensure that devices and circuits connect...

Page 2: ...arness and DSI software For more detailed information on DCM hardware or software refer to the Related Documentation section of this guide Preparing the Environment Ensure that the environment in whic...

Page 3: ...pplication software and device drivers on the host computer in this order NI CompactRIO Device Drivers 16 0 or later Software Calibration Management SCM Toolkit 2016 or later evaluation license Note T...

Page 4: ...ater NI Powertrain Controls DCM Custom Device for NI VeriStand For minimum software support information visit ni com info and enter the Info Code swsupport Hardware Installation Mounting the DCM The D...

Page 5: ...ins all connections required to begin using the DCM with one solenoid injector and one piezo injector The DCM Starter Harness should connect with the devices as shown in Figure 1 What to Use DCM Start...

Page 6: ...ystems The PDU ships configured for 12 V systems by default Complete the following steps to configure the PDU for 24 V systems See Figure 2 below for illustration of fuse placement for 12 V and 24 V s...

Page 7: ...nch top applications which require less than 15 A of current The heavy duty leads 8 AWG included in the DCM Starter Harness kit are recommended for all other applications What to Use Starter harness l...

Page 8: ...T 5 or better shielded twisted pair Ethernet cable Caution To prevent data loss and to maintain the integrity of your Ethernet installation do not use a cable longer than 100 m 328 ft Note The DCM att...

Page 9: ...ur system in the Hostname field 3 Click the Security Configuration icon 4 Click Login 5 In the Login dialog box enter the username admin and leave the password field blank 6 Click OK 7 Click Change Pa...

Page 10: ...re Wizard Tip You must log in if you set a system password c Select the recommended software set for your LabVIEW and NI RIO Device Drivers versions d Click Next e Verify that the summary of software...

Page 11: ...on ni com the software on your DCM may need to be updated Refer to the NI DCM 23XX DSI User s Guide for more information Troubleshooting the DCM Safe Mode and IP Reset The following figure shows the...

Page 12: ...rmation about installing software on the chassis Amber 3 Flashes The chassis is in safe mode Refer to the Measurement Automation Explorer Help for information about safe mode Amber 4 Flashes The softw...

Page 13: ...lp Powertrain Controls Device Drivers User s Manual in LabVIEW to view the Device Drivers User s Manual Browse the DCM section in the Contents tab for information about the DCM API VIs on the FPGA Rea...

Page 14: ...the European Communities using the manufacturer s declaration of conformity This system affords the user protection for electromagnetic compatibility EMC and product safety You can obtain the DoC for...

Page 15: ......

Page 16: ...m patents You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com leg...

Reviews: