NI 9482
CHa-to-CHb...................................... 250 V
rms
max,
Measurement Category II
Isolation
Channel-to-channel
Continuous ........................... 250 V
rms
Withstand ............................. 1,400 V
rms
, verified by a 5 s
dielectric withstand test
Channel-to-earth ground
Continuous ........................... 250 V
rms
Withstand ............................. 2,300 V
rms
, verified by a 5 s
dielectric withstand test
Measurement Category II is for measurements performed on
circuits directly connected to the electrical distribution system.
This category refers to local-level electrical distribution, such as
that provided by a standard wall outlet, for example, 115 V for U.S.
or 230 V for Europe.
Caution
Do
not
connect the NI 9482 to signals or use
for measurements within Measurement Categories III
or IV.