NI 9265 Operating Instructions and Specifications
F
u
ll-scale o
u
tp
u
t c
u
rrent
Minim
u
m.................................... 20.1 mA
Typical ........................................ 20.6 mA
Maxim
u
m ................................... 21.0 mA
O
u
tp
u
t range ..................................... 0 to 20 mA
Compliance voltage .......................... 12 VDC max
Maxim
u
m load.................................. 600
Ω
Acc
u
racy
Measurement Conditions
Percent of
Reading
(Gain Error)
Percent of
Range
*
(Offset Error)
Calibrated, typ (25 °C, ±5 °C)
0.11%
0.19%
Calibrated, max (–40 to 70 °C)
0.25%
0.4%
Uncalibrated, typ (25 °C, ±5 °C)
0.35%
1.4%
Uncalibrated, max (–40 to 70 °C)
0.85%
2.5%
*
Range eq