Summary of Contents for NI-9252

Page 1: ...NI 9252...

Page 2: ...ication 6 Accuracy under Calibration Conditions 6 Worldwide Support and Services 7 Software Calibrating the NI 9252 requires the installation of NI DAQmx 19 1 or later on the calibration system You ca...

Page 3: ...rt for Visual Studio Reference information for NI DAQmx NET methods and NI DAQmx NET properties key concepts and a C enum to NET enum mapping table Test Equipment The following table lists the equipme...

Page 4: ...8 chassis to the system safety ground a Attach a ring lug to a 14 AWG 1 6 mm wire b Connect the ring lug to the ground terminal on the side of the NI cDAQ 9178 using the ground screw c Attach the othe...

Page 5: ...ble 2 NI 9252 AI Voltage Task Physical Channels Range Terminal Configuration cDAQMod8 AI0 7 10 5 V Differential b Configure the AI voltage channel timing according to the following NI 9252 Task Timing...

Page 6: ...equired to adjust the NI 9252 Accuracy Adjustment Complete the following procedure to adjust the accuracy performance of the NI 9252 1 Set the calibrator to Standby mode STBY 2 Connect the NI 9252 to...

Page 7: ...calibration temperature without making any adjustments by initializing an external calibration setting the C Series calibration temperature and closing the external calibration Reverification Repeat...

Page 8: ...i com services for information about the services NI offers Visit ni com register to register your NI product Product registration facilitates technical support and ensures that you receive important...

Page 9: ...ents EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy a...

Reviews: