CALIBRATION PROCEDURE
NI 9216/9226
8-Channel, 400 S/s Aggregate, 0
Ω
to 400
Ω
/4000
Ω,
PT100 RTD/PT1000 RTD C Series Temperature Input Module
This document contains the verification and adjustment procedures for the NI 9216 and NI 9226.
In this document, the NI 9216 and NI 9226 are inclusively referred to as NI 9216/9226. For more
information about calibration solutions, visit
ni.com/calibration
.
Contents
Software.................................................................................................................................... 1
Documentation.......................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 3
Initial Setup............................................................................................................................... 3
Verification............................................................................................................................... 4
Accuracy Verification....................................................................................................... 4
Adjustments.............................................................................................................................. 6
Accuracy Adjustment ....................................................................................................... 6
EEPROM Update ..................................................................................................................... 7
Reverification ........................................................................................................................... 7
Accuracy Under Calibration Conditions .................................................................................. 7
Worldwide Support and Services ............................................................................................. 8
Software
Calibrating the NI 9216/9226 requires the installation of NI-DAQmx 14.5 or later on the
calibration system. You can download NI-DAQmx from
ni.com/downloads
. NI-DAQmx
supports LabVIEW, LabWindows
™
/CVI
™
, ANSI C, Microsoft Visual C++, Microsoft Visual
Basic .NET, and Microsoft Visual C#. When you install NI-DAQmx you only need to install
support for the application software that you intend to use.