NI 9216/9226 Calibration Procedure

Adjustments

The following performance adjustment procedure describes the sequence of operation required 

to adjust the NI 9216/9226.

Accuracy Adjustment

Complete the following procedure to adjust the accuracy performance of the NI 9216/9226.
1. Set the calibrator to Standby mode (STBY).
2. Connect the desired channel of the NI 9216/9226 to the calibrator as shown in Figure 1. 

Connect only one channel at a time.

3. Adjust the desired channel of the NI 9216/9226.

a. Initialize a calibration session on the NI 9216/9226. The default password is 

NI

.

b. Input the external temperature in degrees Celsius.
c. Call the DAQmx Get NI 9216/9226 Calibration Adjustment Points function to obtain 

an array of recommended calibration resistance values.

d. Set the calibrator to a reference value determined by the array of recommended 

calibration voltages.

e. Set calibrator COMP to 4-wire.
f.

Set the calibrator to Operate mode (OPR).

g. Call and configure the NI 9216/9226 adjustment function according to Table 5.

Note

Use appropriate channel number in place of “x” in “AIx.” For example, use 

AI3 for channel 3.

Note

The module acquisition is started each time the adjust function is called.

h. Set the calibrator to Standby mode (STBY).
i.

Repeat steps d through h for each point in the array.

j.

Close the calibration session with action commit.

4. Disconnect the NI 9216/9226 channel from the calibrator.
5. Repeat steps 2 through 4 to calibrate any other desired channel of NI 9216/9226.

Table 5.  

Gain Adjustment Configuration

Physical Channel

Reference Value

AIx

The reference resistance value from the array 

of adjustment points

Summary of Contents for NI 9226

Page 1: ...NI 9226...

Page 2: ...est Equipment 3 Test Conditions 3 Initial Setup 3 Verification 4 Accuracy Verification 4 Adjustments 6 Accuracy Adjustment 6 EEPROM Update 7 Reverification 7 Accuracy Under Calibration Conditions 7 Wo...

Page 3: ...NI 9216 circuitry and specifications NI 9226 Getting Started Guide NI 9226 connection information NI 9226 Datasheet NI 9226 circuitry and specifications NI DAQmx Readme Operating system and applicati...

Page 4: ...midity below 80 Allow a warm up time of at least 10 minutes to ensure that the NI 9216 9226 measurement circuitry is at a stable operating temperature Initial Setup Complete the following steps to set...

Page 5: ...llowing procedure to determine the As Found status of the NI 9216 9226 1 Set the calibrator to Standby mode STBY 2 Connect the desired channel of the NI 9216 9226 to the calibrator as shown in Figure...

Page 6: ...x in AIx For example use AI3 for channel 3 c Configure the task timing according to Table 4 d Start the task e Read the samples from the NI 9216 9226 f Average the samples g Stop and clear the task 7...

Page 7: ...ain an array of recommended calibration resistance values d Set the calibrator to a reference value determined by the array of recommended calibration voltages e Set calibrator COMP to 4 wire f Set th...

Page 8: ...djustment verify that you have met the Test Conditions before returning your device to NI Refer to Worldwide Support and Services for assistance in returning the device to NI Accuracy Under Calibratio...

Page 9: ...ined in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52 227 14 DFAR 252 227 7014 and DFAR 252 227 7015 Wo...

Reviews: