NI 9222/9223 Calibration Procedure
Documentation
Consult the following documents for information about the NI 9222/9223, NI-DAQmx, and
your application software. All documents are available on
and help files install with the
software.
Test Equipment
Table 1 lists the equipment recommended for the performance verification and adjustment
procedures. If the recommended equipment is not available, select a substitute using the
requirements listed in Table 1.
NI cDAQ-9174/9178 USB Chassis Quick Start
NI-DAQmx installation and hardware setup
NI 9222/9223 Operating Instructions and Specifications
NI 9222/9223 specific information, specifications, and calibration interval
NI-DAQmx Readme
Operating system and application software support in NI-DAQmx
LabVIEW Help
LabVIEW programming concepts and reference information about NI-DAQmx
VIs and functions
NI-DAQmx C Reference Help
Reference information for NI-DAQmx C functions and NI-DAQmx C properties
NI-DAQmx .NET Help Support for Visual Studio
Reference information for NI-DAQmx .NET methods and NI-DAQmx .NET
properties, key concepts, and a C enum to .NET enum mapping table
Table 1.
Recommended Equipment
Equipment
Recommended Model
Requirements
Calibrator
Fluke 5700A
Use a high-precision voltage source with an
accuracy
≤
15 ppm and an output impedance
≤
50
Ω
.
Chassis
NI cDAQ-9178
—