NI 9219 Calibration Procedure
|
© National Instruments Corporation
|
39
Note
For operational specifications, refer to the most recent
NI 9219 Operating
Instructions and Specifications
online at
ni.com/manuals
.
Voltage ±4 V
–4 V
4 V
1629
147
Voltage ±1 V
–1 V
1 V
545
14
Voltage ±125 mV
–125 mV
125 mV
528
82
Thermocouple
–125 mV
125 mV
308
82
Current
–25 mA
25 mA
1067
24
4-Wire Resistance 10 k
Ω
0
Ω
10.5 k
Ω
776
61
4-Wire Resistance 1 k
Ω
0
Ω
1.05 k
Ω
771
474
4-Wire RTD Pt1000
0
Ω
5.05 k
Ω
774
122
4-Wire RTD Pt100
0
Ω
505
Ω
764
947
3-Wire RTD Pt1000
0
Ω
5.05 k
Ω
807
122
3-Wire RTD Pt100
0
Ω
505
Ω
797
951
2-Wire Resistance 10 k
Ω
0
Ω
10.5 k
Ω
776
61
2-Wire Resistance 1 k
Ω
0
Ω
1.05 k
Ω
771
475
Quarter-Bridge 350
Ω
0
Ω
390
Ω
760
956
Quarter-Bridge 120
Ω
0
Ω
150
Ω
734
1612
Half-Bridge 500 mV/V
–500 mV/V
500 mV/V
93
205
Full-Bridge 62.5 mV/V
–62.5 mV/V
62.5 mV/V
111
197
Full-Bridge 7.8 mV/V
–7.8125 mV/V
7.8125 mV/V
120
1526
Table 53.
NI 9219 Accuracy Under Calibration Conditions (Continued)
Mode
Range
PPM of
Reading
PPM of
Range
Minimum
Maximum