NI 9218 with LEMO Safety Voltages
Connect only voltages that are within the following limits:
Maximum voltage,
from any pin to any pin
on a single connector
1
±30 V
Isolation
Channel-to-channel, channel-to-Vsup, channel-to-earth,
Vsup-to-earth (up to 5,000 m)
2
Continuous
60 VDC, Measurement
Category I
Withstand
1,000 Vrms, verified by a 5 s
dielectric withstand test
1
The maximum voltage between pin 2 and pin 3 on a single connector is -20 V to
+30 V.
2
Must use crimp contact LEMO plug (784162-01) to maintain these ratings.