±16 V Connections
EX+
VEX+
VEX-
V
SUP+
V
SUP
-
NI 9218
+
-
AI+
AI
-
/EX
-
The NI 9218 provides optional 12 V sensor excitation. To use the
12 V excitation, connect a 9 VDC to 30 VDC power supply to
Vsup, connect the excitation terminals on your sensor to EX+/
EX-, and enable 12 V excitation in your software.
Related Information
NI 9982 ±16 V Connection Pinout
on page 35
NI 9218 Getting Started Guide
|