NI 6601/6602 Calibration Procedure
12
ni.com
6.
Configure
source
of counter 1 to be PFI line 35.
7.
Set the pulse delay for the single-pulse generation.
Traditional NI-DAQ Function Call
LabVIEW Block Diagram
Call
GPCTR_Change_Parameter
with
the following parameters:
deviceNumber
: The value set by MAX
gpctrNum
:
ND_COUNTER_1
paramID
:
ND_SOURCE
paramValue
: ND_PFI_35
Traditional NI-DAQ Function Call
LabVIEW Block Diagram
Call
GPCTR_Change_Parameter
with
the following parameters:
deviceNumber
: The value set by MAX
gpctrNum
:
ND_COUNTER_1
paramID
:
ND_COUNT_1
paramValue
: 2