Appendix A
Specifications
A-2
©
National Instruments Corporation
N
–100
0.68
0.84
1.08
0.03
0.26
0
0.54
0.67
0.86
400
0.42
0.51
0.65
1300
0.57
0.66
0.80
E
–100
0.55
0.62
0.74
0.02
0.28
0
0.41
0.46
0.55
500
0.35
0.40
0.46
1000
0.46
0.50
0.57
T
–150
0.81
0.96
1.17
0.03
0.36
0
0.46
0.55
0.68
400
0.33
0.39
0.47
R
250
0.82
1.16
1.65
0.06
0.12
1000
0.72
0.99
1.37
1767
0.91
1.19
1.60
S
250
0.91
1.28
1.83
0.07
0.13
1000
0.77
1.05
1.47
1767
0.96
1.27
1.72
B
600
1.08
1.64
2.47
0.11
0.00
1000
0.76
1.14
1.69
1820
0.74
1.05
1.50
† Add when thermocouple accessory and NI 4350 is outside 15
°
–35
°
C temperature range
* Add when thermocouple accessory is outside 15
°
–35
°
C temperature range
Error (° C)
15°–35° C, 1 Year
0°–15° C,
35°–55° C
Filter Setting
Temperature
Coefficient
° C/° C
†
Accessory
Error
° C
*
Thermocouple Type
° C
10 Hz
50 Hz
60 Hz