Chapter 3
Signal Connections
© National Instruments Corporation
3-19
Lab-PC+ User Manual
Mode 1 Output Timing
The timing specifications for an output transfer in Mode 1 are as follows:
WR*
OBF*
INTR
ACK*
DATA
T1
T2
T3
T4
T5
T6
Name
Description
Minimum
Maximum
T1
WR* = 0 to INTR = 0
–
450
T2
WR* = 1 to output
–
350
T3
WR* = 1 to OBF* = 0
–
650
T4
ACK* = 0 to OBF* = 1
–
350
T5
ACK* pulse width
300
–
T6
ACK* = 1 to INTR = 1
–
350
All timing values are in nanoseconds.