©
National Instruments Corp.
13
FP-AI-112 and cFP-AI-112
Shock and Vibration
These specifications apply only to the cFP-AI-112.
NI recommends Compact FieldPoint if your application
is subject to shock and vibration.
Operating vibration, random
(IEC 60068-2-64).............................. 10–500 Hz, 5 g
rms
Operating vibration, sinusoidal
(IEC 60068-2-6)................................ 10–500 Hz, 5 g
Operating shock
(IEC 60068-2-27).............................. 50 g, 3 ms half sine,
18 shocks at 6 orientations;
30 g, 11 ms half sine,
18 shocks at 6 orientations
Safety
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement,
control, and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1
•
CAN/CSA-C22.2 No. 61010-1
Note
For UL, hazardous locations, and other safety
certifications, refer to the product label, or visit
ni.com/certification
, search by model number or
product line, and click the appropriate link in the
Certification column.
Electromagnetic Compatibility
Emissions.......................................... EN 55011 Class A at 10 m;
FCC Part 15A above 1 GHz
Immunity........................................... EN 61326:1997 + A2:2001,
Table 1
CE, C-Tick, and FCC Part 15 (Class A) Compliant
Note
For EMC compliance, operate this device with
shielded cabling.