CALIBRATION PROCEDURE

FD-11601

8-Channel, ±10 V Powered-Sensor Voltage Input Device for FieldDAQ

This document contains information for calibrating the FD-11601. For more information about
calibration, visit 

ni.com/calibration

.

Contents

Software Requirements............................................................................................................. 1
Documentation.......................................................................................................................... 2
Test Equipment..........................................................................................................................2
Test Conditions..........................................................................................................................3
Calibration Procedure............................................................................................................... 3

Initial Setup....................................................................................................................... 4
Verification........................................................................................................................4
Adjustment........................................................................................................................ 7
EEPROM Update..............................................................................................................8
Reverification....................................................................................................................8

Accuracy Under Calibration Conditions...................................................................................8
Signal Connector Pinout........................................................................................................... 9
Revision History..................................................................................................................... 10
Worldwide Support and Services............................................................................................ 10

Software Requirements

Calibrating the FD-11601 requires the installation of NI-DAQmx on the calibration system. NI
recommends using the latest NI-DAQmx driver. Earliest driver support version for calibrating
the FD-11601 is listed in the following table.

Table 1. FD-11601 Driver Support

Driver

Earliest Version Support for Device Calibration

 

 

NI-DAQmx

19.0

You can download NI-DAQmx from 

ni.com/downloads

. NI-DAQmx supports many

programming languages, including LabVIEW, LabWindows

/CVI

, C/C++, C#, and Visual

Basic .NET. When you install NI-DAQmx, you only need to install support for the application
software that you intend to use.

Summary of Contents for FD-11601

Page 1: ...9 Revision History 10 Worldwide Support and Services 10 Software Requirements Calibrating the FD 11601 requires the installation of NI DAQmx on the calibration system NI recommends using the latest NI...

Page 2: ...ovides detailed specifications for your FieldDAQ device NI DAQmx Readme Provides operating system and application software support in NI DAQmx NI DAQmx Help Provides information about creating applica...

Page 3: ...d wires act as antennas picking up extra noise that can affect measurements Verify that all connections to the device are secure Use shielded copper wire for all cable connections to the device Use tw...

Page 4: ...ng the Device in MAX When the FieldDAQ device is connected to a network multiple users can access the device To perform any DAQ functionality on the device including reset and self test you must reser...

Page 5: ...alue determined by the array of recommended calibration voltages indicated in the following table Table 3 FD 11601 Test Limits Test Point 1 Year Limits Location Value V Lower Limit V Upper Limit V Neg...

Page 6: ...samples e Average the readings f Clear the task 6 Compare the average to the limits in Table 3 7 Repeat steps 3 through 6 for each test point 8 Set the calibrator to Standby mode STBY 9 Disconnect the...

Page 7: ...n degrees Celsius using the Set Temperature FieldDAQ function b Call the Get FD 11601 Calibration Adjustment Points function c Set the calibrator to a reference value determined by the array of adjust...

Page 8: ...tialize External Calibration function to start a calibration session on the FieldDAQ device The default password is NI 2 Call the DAQmx Set Temperature FieldDAQ Calibration function to input the exter...

Page 9: ...conditions described in this document and are not to be construed as general operational specifications of the FD 11601 For operational specifications refer to the most recent FD 11601 Specifications...

Page 10: ...and Logo Guidelines at ni com trademarks for information on NI trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents cove...

Reviews: