background image

Table 2. Recommended Test Equipment

Equipment

Recommended Model

Minimum Requirements

 

 

 

Calibrator

Fluke 5502A

High-precision voltage source with
gain uncertainty of ≤50 ppm and
offset uncertainty of ≤50 μV

SHM125M-Pigtail I/O
Cable (0.3 m, 1 m, 3 m
lengths) (x4)

NI part number
785950-0R3/01/03

Banana Plug (optional)
(x2)

Use banana plugs with an opening
large enough to accommodate four
22 AWG signal wires

Test Conditions

The following setup and environmental conditions are required to ensure the FD-11601 meets
published specifications:

Keep connections to the device as short as possible. Long cables and wires act as
antennas, picking up extra noise that can affect measurements.

Verify that all connections to the device are secure.

Use shielded copper wire for all cable connections to the device. Use twisted-pairs wire
to eliminate noise and thermal offsets.

Maintain an ambient temperature of 23 ±5 °C. The device temperature will be greater
than the ambient temperature.

Keep relative humidity below 80%.

Allow a warm-up time of at least 10 minutes to ensure that the FieldDAQ device
measurement circuitry is at a stable operating temperature.

Calibration Procedure

The calibration process includes the following steps:
1.

Initial Setup

 on page 4

2.

Verification

 on page 4

3.

Adjustment

 on page 7

4.

EEPROM Update

 on page 8

5.

Reverification

 on page 8

FD-11601 Calibration Procedure

  | 

© National Instruments

  | 

3

Summary of Contents for FD-11601

Page 1: ...9 Revision History 10 Worldwide Support and Services 10 Software Requirements Calibrating the FD 11601 requires the installation of NI DAQmx on the calibration system NI recommends using the latest NI...

Page 2: ...ovides detailed specifications for your FieldDAQ device NI DAQmx Readme Provides operating system and application software support in NI DAQmx NI DAQmx Help Provides information about creating applica...

Page 3: ...d wires act as antennas picking up extra noise that can affect measurements Verify that all connections to the device are secure Use shielded copper wire for all cable connections to the device Use tw...

Page 4: ...ng the Device in MAX When the FieldDAQ device is connected to a network multiple users can access the device To perform any DAQ functionality on the device including reset and self test you must reser...

Page 5: ...alue determined by the array of recommended calibration voltages indicated in the following table Table 3 FD 11601 Test Limits Test Point 1 Year Limits Location Value V Lower Limit V Upper Limit V Neg...

Page 6: ...samples e Average the readings f Clear the task 6 Compare the average to the limits in Table 3 7 Repeat steps 3 through 6 for each test point 8 Set the calibrator to Standby mode STBY 9 Disconnect the...

Page 7: ...n degrees Celsius using the Set Temperature FieldDAQ function b Call the Get FD 11601 Calibration Adjustment Points function c Set the calibrator to a reference value determined by the array of adjust...

Page 8: ...tialize External Calibration function to start a calibration session on the FieldDAQ device The default password is NI 2 Call the DAQmx Set Temperature FieldDAQ Calibration function to input the exter...

Page 9: ...conditions described in this document and are not to be construed as general operational specifications of the FD 11601 For operational specifications refer to the most recent FD 11601 Specifications...

Page 10: ...and Logo Guidelines at ni com trademarks for information on NI trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents cove...

Reviews: