Adjustment

Complete the following procedure to adjust the voltage accuracy of the FieldDAQ device.
1.

Set the calibrator to Standby mode (STBY).

2.

Connect the calibrator to Bank 1 of the FieldDAQ device, as shown in the following
figure. Refer to 

Signal Connector Pinout

 for signal connection information.

Figure 4. Bank 1 Adjustment Connections

AI 0+
AI 0–

HI

LO

AI 1+
AI 1–

Calibrator

FD-11601

AI 2+
AI 2–

AI 3+
AI 3–

Bank 1

2

4

2

4

2

4

2

4

Tie AI 0+, AI 1+, AI 2+, and AI 3+ together.

Tie AI 0-, AI 1-, AI 2-, and AI 3- together.

3.

Initialize a calibration session on the FieldDAQ device. The default password is 

NI

.

4.

Adjust the FieldDAQ device voltage.
a.

Input the external temperature in degrees Celsius using the Set Temperature
FieldDAQ function.

b.

Call the Get FD-11601 Calibration Adjustment Points function.

c.

Set the calibrator to a reference value determined by the array of adjustment points.

d.

Set the calibrator to Operate mode (OPR).

e.

Call and configure the Adjust FD-11601 Calibration function according to the
following table.

Table 6. Adjustment Configuration

Physical Channel

Reference Value

 

 

FD11601-Bank1/ai0:3

A reference value from the array of adjustment points

f.

Repeat steps d through e for each calibration voltage in the array.

5.

Set the calibrator to Standby mode (STBY).

6.

Disconnect the calibrator from the FieldDAQ device.

7.

Connect the calibrator to Bank 2 of the FD-11601, as shown in the following figure.
Refer to 

Signal Connector Pinout

 for signal connection information.

FD-11601 Calibration Procedure

  | 

© National Instruments

  | 

7

Summary of Contents for FD-11601

Page 1: ...9 Revision History 10 Worldwide Support and Services 10 Software Requirements Calibrating the FD 11601 requires the installation of NI DAQmx on the calibration system NI recommends using the latest NI...

Page 2: ...ovides detailed specifications for your FieldDAQ device NI DAQmx Readme Provides operating system and application software support in NI DAQmx NI DAQmx Help Provides information about creating applica...

Page 3: ...d wires act as antennas picking up extra noise that can affect measurements Verify that all connections to the device are secure Use shielded copper wire for all cable connections to the device Use tw...

Page 4: ...ng the Device in MAX When the FieldDAQ device is connected to a network multiple users can access the device To perform any DAQ functionality on the device including reset and self test you must reser...

Page 5: ...alue determined by the array of recommended calibration voltages indicated in the following table Table 3 FD 11601 Test Limits Test Point 1 Year Limits Location Value V Lower Limit V Upper Limit V Neg...

Page 6: ...samples e Average the readings f Clear the task 6 Compare the average to the limits in Table 3 7 Repeat steps 3 through 6 for each test point 8 Set the calibrator to Standby mode STBY 9 Disconnect the...

Page 7: ...n degrees Celsius using the Set Temperature FieldDAQ function b Call the Get FD 11601 Calibration Adjustment Points function c Set the calibrator to a reference value determined by the array of adjust...

Page 8: ...tialize External Calibration function to start a calibration session on the FieldDAQ device The default password is NI 2 Call the DAQmx Set Temperature FieldDAQ Calibration function to input the exter...

Page 9: ...conditions described in this document and are not to be construed as general operational specifications of the FD 11601 For operational specifications refer to the most recent FD 11601 Specifications...

Page 10: ...and Logo Guidelines at ni com trademarks for information on NI trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents cove...

Reviews: