![National Instruments BNC-2090 User Manual Download Page 40](http://html1.mh-extra.com/html/national-instruments/bnc-2090/bnc-2090_user-manual_3370870040.webp)
National Instruments Corporation
B-1
BNC-2090 User Manual
Using the PC-LPM-16
Appendix
B
This appendix contains information on using the PC-LPM-16 with the
BNC-2090.
When using the PC-LPM-16 with the BNC-2090, the functions of the
front panel switches and connectors are different because of the
difference between the MIO I/O connector signals and the PC-LPM-16
I/O connector signals.
Analog I/O
The analog input section of a PC-LPM-16 DAQ board consists of
16 ground-referenced single-ended channels. Therefore, the only valid
configuration for the BNC-2090 is also RSE mode, which uses all
16 analog input BNC connectors. Switches S1 through S8 must be in
the SE position and S9 should be in the RSE position, as shown in
Figure B-1.
Figure B-1.
Switch Configurations for RSE Mode (PC-LPM-16)
Note:
All the switches are required to be in the same position; that is, S1 through
S8 must all be in the SE position and S9 must be in the RSE position to use
the BNC-2090 with the PC-LPM-16.
DAC0OUT and DAC1OUT BNC connectors are for use with the MIO
boards only. These connectors are connected to
±
12 V when the
PC-LPM-16 is in use.
DAC1OUT
DAC0OUT
DIFF
DIFF
SE
ACH9
ACH1
ACH8
DIFF
ACH0
SE
DIFF
ACH10
SE
ACH2
ACH12
ACH4
ACH11
DIFF
ACH3
SE
SE
DIFF
ACH13
SE
ACH5
ACH14
DIFF
DIFF
ACH6
SE
SE
ACH15
N
S
E
R
ACH7
E
R
S
— switch in SE or RSE mode