NI 9269 Calibration Procedure
Figure 1.
Accuracy Verification Connections
3.
Set the DMM to read DC voltage in the minimum range required to cover the NI 9269
output range of ±10 V.
4.
Acquire a sample.
a.
Create and configure an AO voltage channel according to Table 2.
b.
Start the task.
c.
Generate a voltage output according to Table 3.
d.
Wait the appropriate amount of time for the DMM measurement to settle.
e.
Read the NI 9269 output voltage measurement from the DMM.
f.
Clear the task.
5.
Compare the DMM measurement to the test limits in Table 4.
6.
Repeat steps 3 through 5 for each test point in Table 4.
7.
Disconnect the DMM from the NI 9269.
8.
Repeat steps 1 through 7 for each channel on the NI 9269.
Table 2.
NI 9269 AO Voltage Channel Configuration
Input Range
Scaled Units
Terminal Configuration
Min
Max
-10
10
Volts
Differential
Table 3.
NI 9269 Output Configuration
Samples Per Channel
Timeout
1
10.0
NI 9269
DMM
AO
x
–
AO
x
+
HI
LO