NI 9269 Calibration Procedure

|

© National Instruments

|

5

Table 4.  

NI 9269 Accuracy Verification Test Limits

Note

Test limits in Table 4 are derived using the values in Table 6.

Adjustment

The following performance adjustment procedure

 

describes the sequence of operation required 

to adjust the NI 9269.

Accuracy Adjustment

Complete the following procedure to adjust the accuracy performance of the NI 9269.
1.

Adjust the NI 9269.
a.

Initialize a calibration session on the NI 9269. The default password is 

NI

.

b.

Input the external temperature in degrees Celsius.

c.

Call the NI 9269 get C Series adjustment points function to obtain an array of 
recommended calibration voltages for the NI 9269.

d.

Connect the DMM to the NI 9269 as shown in Figure 1.

e.

Set the DMM to read DC voltage in the minimum range required to cover the NI 9269 
output range of ±10 V.

f.

Call and configure the NI 9269 setup calibration function with the DAC value 
obtained from the array of recommended calibration voltages.

g.

Wait the appropriate amount of time for the DMM measurement to settle.

h.

Read the NI 9269 output voltage measurement from the DMM.

i.

Call and configure the NI 9269 adjustment function according to Table 5.

j.

Repeat steps f to i for each calibration voltage in the array.

k.

Close the calibration session.

l.

Disconnect the DMM from the NI 9269.

2.

Repeat step 1 for each channel on the NI 9269. 

Range (V)

Test Point

1-Year Limits

Min

Max

Location

Value (V)

Lower 

Limit (V)

Upper 

Limit (V)

-10.000

10.000

Max

9.500000

9.492730

9.507271

-10.000

10.000

Mid

0.000000

-0.001571

0.001571

-10.000

10.000

Min

-9.500000

-9.507271

-9.492730

Table 5.  

Adjustment Configuration

Physical Channel

Reference Value

cDAQMod8/ao

x

The NI 9269 output voltage measured from the DMM.

Summary of Contents for 9269

Page 1: ...rification 3 Accuracy Verification 3 Adjustment 5 Accuracy Adjustment 5 EEPROM Update 6 Reverification 6 Accuracy Under Calibration Conditions 6 Where to Go for Support 7 Software Calibrating the NI 9269 requires the installation of NI DAQmx 9 6 or later on the calibration system You can download NI DAQmx from ni com downloads NI DAQmx supports LabVIEW LabWindows CVI ANSI C and NET When you instal...

Page 2: ... Operating Instructions and Specifications NI 9269 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW programming concepts and reference information about NI DAQmx VIs and functions NI DAQmx C Reference Help Reference information for NI DAQmx C functions and NI DAQmx C properties NI DAQmx N...

Page 3: ...ting temperature Initial Setup Complete the following steps to set up the NI 9269 1 Install NI DAQmx 2 Make sure that the NI cDAQ 9178 power source is not connected 3 Insert the module into slot 8 of the NI cDAQ 9178 chassis Leave slots 1 through 7 of the NI cDAQ 9178 chassis empty 4 Connect the NI cDAQ 9178 chassis to your host computer 5 Connect the power source to the NI cDAQ 9178 chassis 6 Lau...

Page 4: ...ount of time for the DMM measurement to settle e Read the NI 9269 output voltage measurement from the DMM f Clear the task 5 Compare the DMM measurement to the test limits in Table 4 6 Repeat steps 3 through 5 for each test point in Table 4 7 Disconnect the DMM from the NI 9269 8 Repeat steps 1 through 7 for each channel on the NI 9269 Table 2 NI 9269 AO Voltage Channel Configuration Input Range S...

Page 5: ...DC voltage in the minimum range required to cover the NI 9269 output range of 10 V f Call and configure the NI 9269 setup calibration function with the DAC value obtained from the array of recommended calibration voltages g Wait the appropriate amount of time for the DMM measurement to settle h Read the NI 9269 output voltage measurement from the DMM i Call and configure the NI 9269 adjustment fun...

Page 6: ...tment verify that you have met the Test Conditions before returning your device to NI Refer to Where to Go for Support for assistance in returning the device to NI Accuracy Under Calibration Conditions The values in the following table are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy table is valid for calibration under the following condit...

Page 7: ...e compliance policy and how to obtain relevant HTS codes ECCNs and other import export data Where to Go for Support The National Instruments Web site is your complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Application Engineers National Instruments corp...

Reviews: