NI 9235/9236 Calibration Procedure
6
ni.com
Connecting the Calibrator to the Module
Complete the following steps to connect the calibrator to the module for
both the measurement accuracy verification and the shunt calibration
verification:
1.
Set the calibrator to standby mode (STBY).
2.
Connect the AUX LO terminal of the calibrator to the LO terminal of
the calibrator.
3.
Using discrete connections, connect the LO terminal of the calibrator
to the AI and RC terminals of the module channel you want to verify.
Leave all other channels unconnected.
Note
The three connections to the calibrator LO terminal must be connected at the
calibrator output jack.
Note
Ensure that the connection resistance between the calibrator LO terminal and the
module RC terminal meets the requirement listed in Table 1.
4.
Connect the HI terminal of the calibrator to the EXC0 terminal of the
module.
5.
Connect the AUX HI terminal of the calibrator to the EXC1 terminal
of the module.
Note
You do not have to change the EXC terminal connections for each channel you
verify, because the EXC terminals all connect to a common excitation supply inside the
NI 9235/9236.