NI 9235/9236 Calibration Procedure
10
ni.com
7.
Configure the timing properties for the acquisition using the DAQmx
Timing VI, as shown in the following table.
8.
Turn off the shunt calibration for the channel using the DAQmx
Channel Property Node, as shown in the following table.
9.
Start the acquisition using the DAQmx Start Task VI, as shown in the
following table.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxCfgSampClkTiming
with the following parameters:
taskHandle
:
ta
s
kHandle
source
:
NULL
rate
:
5000
sampleMode
:
DAQmx_Val_FiniteSamp
s
sampsPerChan
:
5000
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxSetAIBridgeShuntCal
Enable
with the following
parameters:
taskHandle
:
ta
s
kHandle
channel
:
NULL
data
:
FALSE
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxStartTa
s
k
with the
following parameter:
taskHandle
:
ta
s
kHandle