Chapter 4
Signal Connections
©
National Instruments Corporation
4-29
A/D conversions generated by either an internal or external CONVERT*
signal are inhibited unless they occur within a DAQ sequence. Scans
occurring within a DAQ sequence may be gated by either the hardware
(AIGATE) signal or software command register gate.
AIGATE Signal
Any PFI pin can externally input the AIGATE signal, which is not
available as an output on the I/O connector. The AIGATE signal can
mask off scans in a DAQ sequence. You can configure the PFI pin you
select as the source for the AIGATE signal in either the level-detection or
edge-detection mode. You can configure the polarity selection for the
PFI pin for either active high or active low.
In the level-detection mode if AIGATE is active, the STARTSCAN signal
is masked off and no scans can occur. In the edge-detection mode, the first
active edge disables the STARTSCAN signal, and the second active edge
enables STARTSCAN.
The AIGATE signal can neither stop a scan in progress nor continue a
previously gated-off scan; in other words, once a scan has started, AIGATE
does not gate off conversions until the beginning of the next scan and,
conversely, if conversions are being gated off, AIGATE does not gate them
back on until the beginning of the next scan.
SISOURCE Signal
Any PFI pin can externally input the SISOURCE signal, which is not
available as an output on the I/O connector. The onboard scan interval
counter uses the SISOURCE signal as a clock to time the generation of the
STARTSCAN signal. You must configure the PFI pin you select as the
source for the SISOURCE signal in the level-detection mode. You can
configure the polarity selection for the PFI pin for either active high or
active low.
The maximum allowed frequency is 20 MHz, with a minimum pulse width
of 23 ns high or low. There is no minimum frequency limitation.