NI 5412 Calibration Procedure
|
© National Instruments
|
17
12. Call
niFgen_SetAttributeViReal64
to set the gain (niFgen property node:
Arbitrary Waveform Output»Arbitrary Waveform Gain
) using the following
parameters:
•
vi
: The session handle returned from
niFgen_init
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_ARB_GAIN
•
value
: The
Gain
value listed in Table 4 for the current iteration.
13. Call
niFgen_SetAttributeViInt32
to choose the positive full-scale DC waveform
(niFgen property node:
Arbitrary Waveform Output»Arbitrary Waveform Handle
)
using the following parameters:
•
vi
: The session handle returned from
niFgen_init
•
channelName
:
"0"
•
attributeID
:
NIFGEN_ATTR_ARB_WAVEFORM_HANDLE
•
value
: The
wfmHandle
from step 8 (positive full-scale handle).
13
0.300000
0.300000
-0.300000
±0.000980
±0.001700
14
0.205000
0.205000
-0.205000
±0.000828
±0.001320
15
0.150000
0.150000
-0.150000
±0.000740
±0.001100
16
0.105000
0.105000
-0.105000
±0.000668
±0.000920
17
0.075000
0.075000
-0.075000
±0.000620
±0.000800
18
0.055000
0.055000
-0.055000
±0.000588
±0.000720
19
0.037500
0.037500
-0.037500
±0.000560
±0.000650
20
0.026000
0.026000
-0.026000
±0.000542
±0.000604
21
0.018500
0.018500
- 0.018500
±0.000530
±0.000574
22
0.013000
0.013000
-0.013000
±0.000521
±0.000552
23
0.009000
0.009000
-0.009000
±0.000514
±0.000536
24
0.006500
0.006500
- 0.006500
±0.000510
±0.000526
Error Positive Full-Scale Value
= (
Measured Positive Full-Scale Value
) - (
Ideal Positive Full-Scale
Value
)
Error Negative Full-Scale Value
= (
Measured Negative Full-Scale Value
) - (
Ideal Negative Full-Scale
Value
)
Table 4.
Values for Verifying the Gain of the Main Analog Path (Continued)
Iteration
Gain
Ideal
Positive
Full-Scale
(Volts)
Ideal
Negative
Full-Scale
(Volts)
Calibration
Test Limit
(Volts)
Published
Specification
(Volts)