NI 5412 Calibration Procedure
|
© National Instruments
|
11
You can verify the following specifications for the NI 5412:
•
Oscillator frequency accuracy
•
DC gain and offset accuracy
•
AC accuracy
•
Flatness
The verification procedure for each of these specifications includes setting up, programming,
and cleaning up.
Note
If any of these tests fail immediately after you perform an external
adjustment, verify that you have met the required test conditions before you return
the NI 5412 to NI for repair.
Refer to Table 2 for information about which instrument to use for verifying each specification.
Refer to Figure 3 for the names and locations of the NI PXI-5412 and the NI PCI-5412 front
panel connectors.
Figure 3.
NI PXI-5412 and NI PCI-5412 Front Panel Connectors