background image

8

13

E. 

If these measures do not unclog the venous system, try using a large (50 cc) syringe to force fluid 

through the tubing.

F. 

If none of these measures work, peel back the skin (soap up arm and skin generously with Ivory® 

liquid detergent) of the arm to the knuckles (do not remove from fingers), and examine all tubing 

for possible kinks. Soap up arm and skin generously with Ivory® liquid detergent, and return skin 

over arm.

CARE OF SIMULATOR

After each class use, disconnect “blood” and flush the venous system. Return synthetic blood to the 

storage bottle. Remove pinch clamp and IV sets from arm. Use tap water to flush venous system and 

wash outside of arm with Ivory® liquid detergent and water. Excess water may be removed from the arm 

by raising the hand, lowering the shoulder, and draining it into a sink or basin. Always remove the pinch 

clamp from shoulder tubing and drain excess water from veins before storing.
Ordinary stains can be removed by washing with soap and warm water. Newsprint, similar printed paper, 

plastic, and ballpoint pen will permanently stain the simulator if prolonged contact occurs. Stubborn 

stains may be removed with Nasco Cleaner (LF09919U) simply by dispensing it on the area and wiping 

with a soft cloth or paper towel.

CAUTIONS

1. 

This synthetic blood is specially formulated to be compatible with the self-sealing veins and plastics 

used in manufacturing the arm.

2. 

DO NOT use dull or burred needles, as these will cause leaks in the system. Burred needles will 

cause permanent damage. Use smaller needles (22-gauge or smaller).

3. 

DO NOT allow “blood” to dry on simulator — it may stain the arm.

4. 

Use only 500 cc of infusion fluid, as a larger amount will also increase the pressure of the venous 

system, resulting in leaks.

5. 

DO NOT clean the simulator with solvents or corrosive material, as they will damage it.

6. 

DO NOT use for subcutaneous injection. Nasco’s Intradermal Injection Simulator (LF01008U) is 

specially designed for intradermal injection training and practice.

7. 

Nasco Vein Tubing Sealant Kit (LF01099U) will extend the useful life of the tubing.

SUPPLIES/REPLACEMENT PARTS FOR CHILD INJECTABLE TRAINING ARM

LF00845U

 

Life

/

form

®

 Venous Blood, 1 quart

LF00846U

 

Life

/

form

®

 Venous Blood, 1 gallon

LF01099U

  Vein Tubing Sealant Kit

LF03629U

  Skin and Vein Replacement Kit

LF09919U

  Nasco Cleaner

Lastly, take the loose end of the tubing and push the 

male luer fitting into it. 

(See figure 17.) 

Attach the 

male luer fitting to the fitting marked CUFF at the 

top of the unit. Once all of the necessary connec-

tions have been properly made, go to the section 

titled “Calibration Procedures” and calibrate the 

unit. When the calibration procedures have been 

completed, the simulator is ready for use.

Figure 15

Figure 16

Figure 17

Figure 14

PREPARING YOUR EQUIPMENT FOR USE WITH THE 

NASCO 

Life

/

form

®

 CHILD BLOOD PRESSURE SIMULATOR

To adapt your sphygmomanometer for use with the simulator, it is first necessary to obtain the luer fitting 

and the T-fitting included with the simulator, which is also available through your Nasco catalog sales 

office. Using a scissors, carefully cut the pressure line of the sphygmomanometer about 2" from the 

gauge. 

(See figure 14.)

Take the T-fitting and insert it between the two ends of tubing that were previously cut. 

(See figure 15.)

 

Assemble the portion of pressure line tubing over the remaining barbed end on the T-fitting. 

(See figure 16.)

SUPPLIES/REPLACEMENT PARTS FOR CHILD BLOOD PRESSURE SIMULATOR

LF01096U

  Electronic Control Unit with Sphygmomanometer

LF01189U

  Amplifier/Speaker System

HC190001INM14.indd   15-16

2/20/19   11:23 AM

Summary of Contents for CRiSis LF03616U

Page 1: ...HEALTHCARE ChildCRiSis Manikin Simulator LF03616U INSTRUCTION MANUAL WARNING Products may contain dry natural rubber 5 Year Warranty erials ...

Page 2: ...2 TABLE OF CONTENTS Introduction 3 Airway Management Trainer 4 IV Arm 5 Blood Pressure Arm 9 Defibrillation Chest Skin 14 Intraosseous Infusion Femoral Access Leg 16 ...

Page 3: ...al access ASSEMBLY OF YOUR CHILD CRiSis MANIKIN To facilitate packing your manikin comes with the arms and legs unattached To connect the arms to the torso place each in a position such that the hand is above the head See figure 1 Align the arm fitting with the torso socket and plug the arm into place Finish by rotating the arm forward so it lies along the side of the body To connect the legs to t...

Page 4: ...te both the oral cavity and the endotracheal tube with the pump spray lubricant provided See figure 1 Note Nasco recommends the use of the provided lubricant or a similar vegetable based lubricant for the Child Trainer The use of a silicone based or similar lubricant may seriously damage the simulator thus voiding Nasco s warranty on the trainer CLEANING THE FACE Normal soil can be removed from th...

Page 5: ...eplacement Kits are available through Nasco see list of supplies LIST OF COMPONENTS 3 cc Syringe with Needle 12 cc Syringe with Needle 2 IV Bags Needle Butterfly 1 Pinch Clamp Small Towel Arm INTERNAL STRUCTURE Internally the vascular structure rubber tubing begins at the shoulder and continues under the arm crosses the antecubital fossa forearm makes a loop in the back of the hand and then return...

Page 6: ...fter filling the venous system according to instructions the arm is now ready to practice drawing blood Blood can be drawn anywhere along the pathway of the vein See figure 1 Distilled water should be used to prepare the sites Synthetic blood will actually be aspirated once the vein is properly punctured 8 Small diameter needles 22 gauge or smaller should be used B Preparing the Arm for Intravenou...

Page 7: ...flow into container B until bubbles in tubing disappear then regulate blood flow from bag A using the clamp System is now full of blood and pressurized Blood can now be drawn anywhere along the pathway of the vein 2 Intravenous infusion Insert the butterfly needle into the lumen of the vein A flashback of blood is proof the needle was inserted correctly Close the clamp on the bag A tubing and disc...

Page 8: ...occurs Stubborn stains may be removed with Nasco Cleaner LF09919U simply by dispensing it on the area and wiping with a soft cloth or paper towel CAUTIONS 1 This synthetic blood is specially formulated to be compatible with the self sealing veins and plastics used in manufacturing the arm 2 DO NOT use dull or burred needles as these will cause leaks in the system Burred needles will cause permanen...

Page 9: ...control box has a battery saving feature which will turn the unit off after about 8 10 minutes if no keys are used within that period of time Figure1 ective set been that the g s AMP e the ce he eing le he perly oma ssible rm ssure mHg at dow s 2 ed at 13 BRA Figure2 On Off Switch The next step is to connect the simulated arm and speaker assembly along with the sphygmomanometer cuff and gauge asse...

Page 10: ...utomatically be set to 0 Figure5 Figure6 Figure7 Figure8 PALPABLE PULSE FEATURE The Nasco Life form Child Blood Pressure Simulator also incorporates a palpable pulse at the radial location See figure 9 Palpations can be felt upon start up of the unit or after blood pressure settings have been made Press the Menu key repeatedly until Set PALPATION menu appears Pulse ON is defaulted and enables the ...

Page 11: ...henomenon Pressing the Gap key simply turns the gap function off or on When the key is pressed a message will briefly appear that the auscultatory gap is enabled or disabled Also the main display will show at the bottom right of the display either AGap Y for on or AGap N for off The arrow up and down keys also control the volume of the sounds that are present in the arm From the main menu press th...

Page 12: ...tion of the measurement sequence LOW BATTERY INDICATOR When the battery supply diminishes to a level near the point that the unit will no longer function properly the low batt segment of the systolic pressure display will activate when the pressure in the sphygmoma nometer cuff reaches above 20 mmHg At this point the batteries should be replaced as soon as possible in order to insure proper operat...

Page 13: ...e form CHILD BLOOD PRESSURE SIMULATOR To adapt your sphygmomanometer for use with the simulator it is first necessary to obtain the luer fitting and the T fitting included with the simulator which is also available through your Nasco catalog sales office Using a scissors carefully cut the pressure line of the sphygmomanometer about 2 from the gauge See figure 14 Take the T fitting and insert it be...

Page 14: ...st skin Energy levels in excess of 360 joules may cause irreparable damage to the chest skin circuitry and patient simulator being used thus voiding Nasco s warranty and endangering your equipment Nasco assumes no liability for damage or injury that may be caused by the use and or misuse of this equipment All normal safety precautions for defibrillation training should be followed and energy level...

Page 15: ... connections on the patient simulator one or more may be disconnected 2 Check to make sure your patient simulator is plugged in and working properly Problem ECG wave is inverted Solution Recheck the position of the red and black lead snaps on the patient simulator Note If the defibrillation chest skin is not functioning or wiring comes undone please contact us to repair or replace the unit Failure...

Page 16: ...1 60 cc Syringe K Y Jelly Wax 1 2 oz Liquid Lubricant 2 Leg Skins 2 Mixing Bottles w Blood Mix IV Bag THE INTRAOSSEOUS INFUSION FEMORAL ACCESS LEG GENERAL INSTRUCTIONS FOR INTRAOSSEOUS INFUSION USE 1 Place the simulator on a flat surface such as a tabletop Lay towels under the knees for support and to absorb any overflow 2 To prepare the blood mixture combine the red coloring 1 full tube of K Y Je...

Page 17: ...ed Extreme caution should be taken to avoid pushing the needle completely through the simulator injuring the person performing the proce dure Nasco cannot be responsible for injuries resulting from improper use of the simulator ure1 4 Connect the end of the tubing to the bone piece Apply liquid lubricant to the entire bone and slide into posi tion in the leg See figure 3 IMPORTANT 5 Make your firs...

Page 18: ...e access site tubing See figure 6 4 Open the clamp on the IV bag 5 To charge the vein make your first needle inser tion in the vein and pull back on the syringe to remove air from the vein Repeat until blood is withdrawn The vein is now charged and subse quent insertions will produce an immediate flow of blood CLEAN UP PROCEDURES 1 Close the clamp on the IV bag pull the fitting from the femoral ac...

Page 19: ...iSis Manikin with Arrhythmia Tutor LF03709U Infant CRiSis Manikin LF03953U CRiSis Manikin Complete LF03955U Deluxe CRiSis Manikin LF03956U Deluxe Plus CRiSis Manikin LF03965U Adult CRiSis Auscultation Manikin LF03966U Adult CRiSis Auscultation Manikin with ECG Simulator LF03709U LF03965U Actual product may vary slightly from photo Nasco reserves the right to change product color materials supplies...

Reviews: