background image

12

9

THE BLOOD PRESSURE ARM

GENERAL INSTRUCTIONS FOR USE INSTALLING THE BATTERIES

Take the Blood Pressure Electronic Control Unit from the box and turn it over, placing it face down onto a 

padded work surface. Locate the “Open” compartment on the back of the panel where the batteries are 

to be installed. 

(See figure 1.) 

Place your thumb or index finger on the “Open” compartment and push up.

This will open the battery compartment. The compartment is marked as to the “+” and “–” positions of 

the batteries. The battery bracket is now accessible to the user. 
Install six “AA” batteries as indicated by the orientation diagram embossed in the bottom of the bracket. 

It is recommended that alkaline batteries be used for increased battery life. After the batteries have 

been properly installed, reassemble the Blood Pressure Simulator by simply reversing the disassembly 

procedures. Place the unit face up on the work surface and turn it on by pressing the on/off switch on 

the top right of the unit. 

(See figure 2.)

 Observe the display and verify that a readable display is present, 

indicating proper battery installation. 

Note: 

The control box has a battery saving feature which will turn 

the unit off after about 8-10 minutes if no keys are used within that period of time.

Figure 1

Compare the results of reading the systolic and diastolic blood pressures on the gauge with the respective 

settings on the simulator. If the readings were accurate, they should compare favorably with the preset 

values. If the auscultatory gap had been selected, then an absence of an audible pulse would have been 

noticed during what would have been the phase 2 Korotkoff sound. It is this lack of an audible pulse that 

is considered an auscultatory gap. Note that in reality the auscultatory gap can be present in either the 

phase 1 or phase 2 Korotkoff sounds. If it is desired to demonstrate the sounds heard while measuring 

a subject’s blood pressure to the trainee or group of trainees, an auxiliary amplifier/speaker system is 

available (LF01189U). If the auxiliary speaker is used, the speaker is plugged into the EXT AUDIO AMP 

jack located next to the ARM output jack at the top of the unit. Adjust the volume control to increase the 

output of the auxiliary speaker amplifier, and proceed with the blood pressure measurement sequence 

as it would normally be performed, except that the stethoscope need not be used. Instead, listen to the 

sounds as they emanate from the speaker amplifier, noting the differences in the Korotkoff phases being 

presented.
The Nasco 

Life

/

form

®

 Child Blood Pressure Simulator is programmed to demonstrate the five Korot-

koff phases, including an auscultatory gap, which can be heard during auscultation of a subject while 

measuring the subject’s blood pressure. Each is distinctly different and present for only a portion of the 

measurement sequence.

LOW BATTERY INDICATOR

When the battery supply diminishes to a level near the point that the unit will no longer function properly, 

the “low batt” segment of the systolic pressure display will activate when the pressure in the sphygmoma-

nometer cuff reaches above 20 mmHg. At this point, the batteries should be replaced as soon as possible 

in order to insure proper operation of the unit. Refer to the section titled “Installing the Batteries.”

CALIBRATION PROCEDURES

To calibrate the simulator, set the unit up as described in the section titled “Using The Nasco 

Life

/

form

®

 

Child Blood Pressure Simulator.” Apply the cuff to the simulated arm. Set the control box systolic pressure 

to 150 mmHg and set the diastolic pressure to 70 mmHg. Proceed with a simulated blood pressure 

measurement. Note the discrepancy in the readings between the gauge and the control box. Set the 

systolic “offset.” 

Example: 

If the blood pressure reading was taken and the sounds started at 148 mmHg, 

then the offset is +2. If the sounds started at 152 mmHg, the offset is –2. For this example, assume that 

the sounds started at 148 mmHg. Press and hold the Calibration key until the systolic correction window 

appears. 

(See figure 12.) 

Using the arrow up key, set the correction to +2. Set the diastolic offset.

Example:

 If the blood pressure reading was taken and the sounds stopped at 72 mmHg, the offset is –2; 

if the sounds stopped at 68 mmHg, the offset is +2. For this example, assume that the sounds stopped at 

72 mmHg. From the systolic window, press the Menu key to change the diastolic window. 

(See figure 13.) 

With the arrow down key, set the correction to –2. Press the Menu key again and the message “CALIBRA-

TION COMPLETE” will appear. The main menu window will be displayed.

Figure 12

Figure 2

On/Off Switch

The next step is to connect the simulated arm and speaker assembly along with the sphygmomanometer 

cuff and gauge assembly included with the unit. First, locate the end of the pressure line attached to the 

sphygmomanometer that has the male luer fitting attached to it. Attach this to the female luer fitting at 

the top of the unit marked CUFF. 

(See figure 3.) 

After the pressure line fitting has been properly installed, 

locate the plug that is at the end of the wire which extends from the simulated arm assembly. Insert the 

plug from the arm into the jack at the top of the unit marked ARM. 

(See figure 4.) 

At this point the Blood 

Pressure Simulator is ready for use. The unit has been factory calibrated for use with the accessories 

included in the kit. No further calibration adjustments should be necessary at this time. If the unit is to be 

used with a sphygmomanometer other than that supplied with the unit, or if recalibration is necessary at 

a later date, then see the section titled “Calibration Procedures.”

Figure 3

Figure 4

Figure 13

HC190001INM14.indd   17-18

2/20/19   11:23 AM

Summary of Contents for CRiSis LF03616U

Page 1: ...HEALTHCARE ChildCRiSis Manikin Simulator LF03616U INSTRUCTION MANUAL WARNING Products may contain dry natural rubber 5 Year Warranty erials ...

Page 2: ...2 TABLE OF CONTENTS Introduction 3 Airway Management Trainer 4 IV Arm 5 Blood Pressure Arm 9 Defibrillation Chest Skin 14 Intraosseous Infusion Femoral Access Leg 16 ...

Page 3: ...al access ASSEMBLY OF YOUR CHILD CRiSis MANIKIN To facilitate packing your manikin comes with the arms and legs unattached To connect the arms to the torso place each in a position such that the hand is above the head See figure 1 Align the arm fitting with the torso socket and plug the arm into place Finish by rotating the arm forward so it lies along the side of the body To connect the legs to t...

Page 4: ...te both the oral cavity and the endotracheal tube with the pump spray lubricant provided See figure 1 Note Nasco recommends the use of the provided lubricant or a similar vegetable based lubricant for the Child Trainer The use of a silicone based or similar lubricant may seriously damage the simulator thus voiding Nasco s warranty on the trainer CLEANING THE FACE Normal soil can be removed from th...

Page 5: ...eplacement Kits are available through Nasco see list of supplies LIST OF COMPONENTS 3 cc Syringe with Needle 12 cc Syringe with Needle 2 IV Bags Needle Butterfly 1 Pinch Clamp Small Towel Arm INTERNAL STRUCTURE Internally the vascular structure rubber tubing begins at the shoulder and continues under the arm crosses the antecubital fossa forearm makes a loop in the back of the hand and then return...

Page 6: ...fter filling the venous system according to instructions the arm is now ready to practice drawing blood Blood can be drawn anywhere along the pathway of the vein See figure 1 Distilled water should be used to prepare the sites Synthetic blood will actually be aspirated once the vein is properly punctured 8 Small diameter needles 22 gauge or smaller should be used B Preparing the Arm for Intravenou...

Page 7: ...flow into container B until bubbles in tubing disappear then regulate blood flow from bag A using the clamp System is now full of blood and pressurized Blood can now be drawn anywhere along the pathway of the vein 2 Intravenous infusion Insert the butterfly needle into the lumen of the vein A flashback of blood is proof the needle was inserted correctly Close the clamp on the bag A tubing and disc...

Page 8: ...occurs Stubborn stains may be removed with Nasco Cleaner LF09919U simply by dispensing it on the area and wiping with a soft cloth or paper towel CAUTIONS 1 This synthetic blood is specially formulated to be compatible with the self sealing veins and plastics used in manufacturing the arm 2 DO NOT use dull or burred needles as these will cause leaks in the system Burred needles will cause permanen...

Page 9: ...control box has a battery saving feature which will turn the unit off after about 8 10 minutes if no keys are used within that period of time Figure1 ective set been that the g s AMP e the ce he eing le he perly oma ssible rm ssure mHg at dow s 2 ed at 13 BRA Figure2 On Off Switch The next step is to connect the simulated arm and speaker assembly along with the sphygmomanometer cuff and gauge asse...

Page 10: ...utomatically be set to 0 Figure5 Figure6 Figure7 Figure8 PALPABLE PULSE FEATURE The Nasco Life form Child Blood Pressure Simulator also incorporates a palpable pulse at the radial location See figure 9 Palpations can be felt upon start up of the unit or after blood pressure settings have been made Press the Menu key repeatedly until Set PALPATION menu appears Pulse ON is defaulted and enables the ...

Page 11: ...henomenon Pressing the Gap key simply turns the gap function off or on When the key is pressed a message will briefly appear that the auscultatory gap is enabled or disabled Also the main display will show at the bottom right of the display either AGap Y for on or AGap N for off The arrow up and down keys also control the volume of the sounds that are present in the arm From the main menu press th...

Page 12: ...tion of the measurement sequence LOW BATTERY INDICATOR When the battery supply diminishes to a level near the point that the unit will no longer function properly the low batt segment of the systolic pressure display will activate when the pressure in the sphygmoma nometer cuff reaches above 20 mmHg At this point the batteries should be replaced as soon as possible in order to insure proper operat...

Page 13: ...e form CHILD BLOOD PRESSURE SIMULATOR To adapt your sphygmomanometer for use with the simulator it is first necessary to obtain the luer fitting and the T fitting included with the simulator which is also available through your Nasco catalog sales office Using a scissors carefully cut the pressure line of the sphygmomanometer about 2 from the gauge See figure 14 Take the T fitting and insert it be...

Page 14: ...st skin Energy levels in excess of 360 joules may cause irreparable damage to the chest skin circuitry and patient simulator being used thus voiding Nasco s warranty and endangering your equipment Nasco assumes no liability for damage or injury that may be caused by the use and or misuse of this equipment All normal safety precautions for defibrillation training should be followed and energy level...

Page 15: ... connections on the patient simulator one or more may be disconnected 2 Check to make sure your patient simulator is plugged in and working properly Problem ECG wave is inverted Solution Recheck the position of the red and black lead snaps on the patient simulator Note If the defibrillation chest skin is not functioning or wiring comes undone please contact us to repair or replace the unit Failure...

Page 16: ...1 60 cc Syringe K Y Jelly Wax 1 2 oz Liquid Lubricant 2 Leg Skins 2 Mixing Bottles w Blood Mix IV Bag THE INTRAOSSEOUS INFUSION FEMORAL ACCESS LEG GENERAL INSTRUCTIONS FOR INTRAOSSEOUS INFUSION USE 1 Place the simulator on a flat surface such as a tabletop Lay towels under the knees for support and to absorb any overflow 2 To prepare the blood mixture combine the red coloring 1 full tube of K Y Je...

Page 17: ...ed Extreme caution should be taken to avoid pushing the needle completely through the simulator injuring the person performing the proce dure Nasco cannot be responsible for injuries resulting from improper use of the simulator ure1 4 Connect the end of the tubing to the bone piece Apply liquid lubricant to the entire bone and slide into posi tion in the leg See figure 3 IMPORTANT 5 Make your firs...

Page 18: ...e access site tubing See figure 6 4 Open the clamp on the IV bag 5 To charge the vein make your first needle inser tion in the vein and pull back on the syringe to remove air from the vein Repeat until blood is withdrawn The vein is now charged and subse quent insertions will produce an immediate flow of blood CLEAN UP PROCEDURES 1 Close the clamp on the IV bag pull the fitting from the femoral ac...

Page 19: ...iSis Manikin with Arrhythmia Tutor LF03709U Infant CRiSis Manikin LF03953U CRiSis Manikin Complete LF03955U Deluxe CRiSis Manikin LF03956U Deluxe Plus CRiSis Manikin LF03965U Adult CRiSis Auscultation Manikin LF03966U Adult CRiSis Auscultation Manikin with ECG Simulator LF03709U LF03965U Actual product may vary slightly from photo Nasco reserves the right to change product color materials supplies...

Reviews: