44
Sp
ec
if
i-
ca
ti
ons
•
The E- and H-field components are
measured simultaneously. The largest
value is displayed.
Audible warning
•
Built-in loudspeaker
•
Audible alarm if alarm threshold exceeded
(50% of relevant standard value)
•
BEEP repetition rate: 2 Hz starting at 50%
of standard, 4 Hz starting at 100% of
standard.
Frequency response evaluation and limit
values according to different standards
ESM “RadMan” is available in a wide range of
different versions. Please refer to the tables on
page 51 and page 52. The frequency range
values printed on the instrument apply in each
case.
Frequency response
Frequency response referred to calibration
frequency with ideal approximated limit curve
Sensor
Frequency range
Typical deviation
H-field
up to 1 GHz
+3 dB / -3 dB
E-field
up to 3 GHz
+3 dB / -3 dB
3 GHz to 10 GHz
+4 dB / -3 dB
10 GHz to 18 GHz
+6 dB / -3 dB
18 GHz to 40 GHz
+6 dB / -10 dB
Specifications
Downloaded from