Preparation for use

 

 

 

 
3.4 Preparation for use 
 
 

 
The male BNC plug of the Pulse Limiter must be connected to the RF input 
socket of the EMI test receiver. 
 

 

 

 

 

 

 

 

Pulse  Limiter  usage  under  some  condition  may  affect  the 
measurement result in respect to test signal characteristics. 
 

 

When  test  signal  has  very  high  pulses  the  Pulse  Limiter  may  reach 
saturation and affect measurement accuracy. 

 

When  very  strong  test  signal  is  suspected  the  usage  of  a  10  dB  or 
more  50 

Ω

  attenuator  in  front  of  the  Pulse  Limiter  is  strongly 

recommended. 

 
A  verify  if  Pulse  Limiter  is  reaching  saturation  can  be  carried  out  simply 
performing  two  measurement  of  the  suspected  strong  signal,  respectively 
with and without a 10 dB, or more, 50 

Ω

 attenuator connected between test 

signal and Pulse Limiter. 
The  measured  test  signal  difference  must  match  the  attenuation 
introduced:  if  the  difference  between  the  two  measured  signals  is  lower 
than  attenuator  value  it  means  that  without  using  attenuator  the  Pulse 
Limiter is reaching saturation, affecting therefore measurement accuracy. 
In  this  case  the  final  measurement  must  be  carried  out  with  attenuator 
inserted taking into account the attenuator value in the test signal reading.

 

 

 

 

 
 

 
Before  connecting  Pulse  Limiter  to  the  associated  test  instrument, 
ensure  that  an  uninterruptible  safety  earth  ground  is  provided  from 
the  main  power  source  to  the  EMI  test  receiver  protective  earth 
connection. 
 
To  avoid  hazardous  electrical  shock,  prior  to  energizing  either  unit 
and  prior  Pulse  Limiter  connection,  verify  that  a  common  ground 
exists between EMI test receiver and the equipment under test. 
 
Any  interruption  or  loosening  of  the  protective  earth  ground 
conductor, either inside or outside the units or in an extension cable 
will  cause  a  potential  shock  hazard  that  could  result  in  personal 
injury. 
 
Verify the safety earth ground functionality before operation. 

 
 
 
 
 

3.5 Equipment cleaning 
 

Use  a  clean,  dry  non  abrasive  cloth  for  external  cleaning  of  the  Pulse 
Limiter. 
 

 

 

 
To  clean  the  equipment  do  not  use  any  solvent,  thinner,  turpentine, 
acid,  acetone  or  similar  matter  to  avoid  damage  to  external  plastic 
and surfaces. 

 

 

         A

WARNING

         A

WARNING

         A

WARNING

         A

WARNING

Summary of Contents for PMM PL-01

Page 1: ...THE INSTRUMENT You can find the Serial Number in a side of the box Serial Number is in the form 0000X00000 The first four digits and the letter are the Serial Number prefix the last five digits are the Serial Number suffix The prefix is the same for identical instruments it changes only when a configuration change is made to the instrument The suffix is different for each instrument Document PL01E...

Page 2: ...ment must be disposed of in accordance with local regulations This product conforms to the WEEE Directive of the European Union 2002 96 EC and belongs to Category 9 Monitoring and Control Instruments You can return the instrument to us free of charge for proper environment friendly disposal You can obtain further information from your local NARDA Sales Partner or by visiting our website at www nar...

Page 3: ...ation 1 2 Introduction to PL01 Pulse Limiter 1 3 Insertion loss 1 4 Environment 1 5 Return for service 2 Main Specification Page 1 1 1 1 1 2 1 2 1 2 Page 2 1 Main Specification 2 1 3 Preparation for use Page 3 1 Introduction 3 1 3 2 Packing unpacking 3 1 3 3 Initial inspection 3 1 3 4 Preparation for use 3 2 3 5 Equipment cleaning 3 2 ...

Page 4: ...IV Contents Figures Figure 1 1 PMM PL01 Typical frequency response Page 1 1 Tables Table 2 1 Main Specification Page 2 1 ...

Page 5: ...t In case of connection of the device to the power mains make sure before connection that the voltage selected on the voltage switch and the fuses are adequate for the voltage of the actual mains Devices in Safety Class I equipped with connection to the power mains by means of cord and plug can only be plugged into a socket equipped with a ground wire Any interruption or loosening of the ground wi...

Page 6: ...o sul Neva SV ITALY based on the following harmonized European Standards successfully applied Safety EN 61010 1 undated reference applies to all editions EMC EN 61326 1 undated reference applies to all editions declares under its sole responsibility that the product PMM PL 01 Pulse Limiter conforms with the essential requirements of the Low Voltage Directive 2014 35 EU of the EMC Directive EMC 201...

Page 7: ...rom artificial mains network LISN the Pulse Limiter PL01 must be used particularly when very strong interfering signal is suspected The precision input attenuator and the input mixer of the receiver are therefore protected from being damaged On the other hand if the disturbance signal and spikes do not exceed the maximum input voltage permissible for the receiver the use of the Pulse Limiter can b...

Page 8: ...ed to NARDA for service please complete the service questionnaire enclosed with the Operating Manual and attach it to the instrument To minimize the repair time be as specific as possible when describing the failure If the failure only occurs under certain conditions explain how to duplicate the failure If possible reuse of the original packaging to ship the equipment is preferable In case other p...

Page 9: ...t temperature must be 0 to 45 TABLE 2 1 Main specifications Electrical characteristics Performance Limits Insertion loss max Input impedance VSWR with 50 Ohm termination Frequency response Max permissible input power RF input RF output Rated temperature Storage temperature 0 4 dB 50 Ohm 1 25 9 kHz 30 MHz 0 5 W BNC female connector BNC male connector 0 to 45 C 25 to 70 C Document PL01EN 00708 NARDA...

Page 10: ...2 2 Main specifications This page has been intentionally left blank ...

Page 11: ...d be kept until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically Verify the accessories availability in the shipping container referring to the accessories check list enclosed with the Operating Manual Notify any damage to the carrier as well as the NARDA Representative 3 3 Initial inspection Inspect the Pulse Limiter...

Page 12: ...miter is reaching saturation affecting therefore measurement accuracy In this case the final measurement must be carried out with attenuator inserted taking into account the attenuator value in the test signal reading Before connecting Pulse Limiter to the associated test instrument ensure that an uninterruptible safety earth ground is provided from the main power source to the EMI test receiver p...

Page 13: ...u would like to submit to the attention of our service organization is kept in great consideration Moreover we are continuously improving our quality but we know this is a never ending process We would be glad if our present efforts are pleasing you Should one of your pieces of NARDA equipment need servicing you can help us serve you more effectively filling out this card and enclosing it with the...

Page 14: ...Suggerimenti Commenti Note Suggestions Comments Note ...

Reviews: