![NanoVibronix PainShield User Manual Download Page 23](http://html2.mh-extra.com/html/nanovibronix/painshield/painshield_user-manual_3242499023.webp)
PainShield® MD User Manual
23
Appendix A:
Electromagnetic Compliance
Table 1: PainShield® MD Electromagnetic Emission
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
PainShield® MD is intended for use in the electromagnetic environment specified below.
The customer or user of PainShield® MD should assure that it is used in such an
environment.
Emission test
Compliance
Electromagnetic Environment – Guidance
RF emissions,
CISPR 11
Group 1
PainShield® MD uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions,
CISPR 11
Class B
PainShield® MD is not directly connected to the
Public Mains Network.
Harmonic emissions,
IEC 61000-3-2
Class B
Voltage fluctuations /
flicker emissions IES
61000-3-3
Complies
Table 2: PainShield® MD Electromagnetic Immunity
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
PainShield® MD is intended for use in the electromagnetic environment specified below.
The customer or user of PainShield® MD should assure that it is used in such an
environment.
Immunity test
IEC 60601-1-2
Test level
Compliance
Level
Electromagnetic Environment –
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast
transfer/burst
IEC 61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±2 kV for
power supply
lines
±1 kV for
input/output
lines
Mains power quality should be
that of a typical commercial
environment.
Summary of Contents for PainShield