PainShield® MD User Manual
20
6.7
Replacing the Actuator
It is recommended to replace the actuator after 30 treatments.
New actuators can be ordered from NanoVibronix or your local distributor.
1.
Switch OFF the PainShield driver (to prevent the driver from sounding
an alarm when it is disconnected from the actuator).
2.
Hold the driver cable socket in one hand and the actuator cable plug
in the other hand and then pull them apart.
CAUTION
Never attempt to disconnect the driver from the actuator by holding
the driver, actuator, or either of the cables, and pulling on them.
3.
Connect the new actuator cable plug into the driver cable socket,
making sure to orient the plug correctly in relation to the socket, and
making sure that the plug is fully inserted and secured.