ADAMII™ ©2016 NanoEnTek Inc.
17
Calibration
Calibration
procedure
12. Select PerCP channel and adjust the focus using the focus knob.
13. Press the ‘Save FL Focus’ after adjusting focus. The focus of other
fluorescence channels will be set automatically.
14. Press the ‘Run’ button to start the auto calibration. Calibration will
take about 3 minutes to count the 24 frames.
☞
Note: Save the focus in ‘Bright’ and ‘PerCP’ channel.
The other channels will be automatically focused.
3 min.
11. Press the ‘Save BR Focus’ after adjusting focus.
•
Bright - Good Focus
: Clear circumference of the beads, Dark and
clear contrast to the background
•
Bright - Bad Focus
: Blurry and unclear bead circumference
•
FL - Good Focus
: Bright fluorescence beads with clear contrast to the
background
•
FL - Bad Focus
: Weak fluorescence beads with unclear contrast to the
background