14
Calibration
Calibration
procedure
6. Attach dry-proof closure tape to prevent drying.
[Attach dry-proof closure tape]
Low volume
Correct volume
Warning:
• Calibration bead loading error
- Ensure that 25μL of calibration bead is fully loaded on the ADAMII™
assay slide.
- The instrument will not detect in low or high loading volumes.
CAUTION:
• Avoid bubbles as they may negatively affect results.
• Boundary between black and white on the tape should be lined up
with the edge of the slide.
Dry-proof closure tape