56

MXA51005

EN | Baby breathing monitor

9. 

OPERATOR WARNING

Event

Light indication

Acoustic indication

Low battery

The indicator light with the battery symbol 

flashes slowly and red 

(100 ms light/2 s pause).

No acoustic signal

Critically low battery

The indicator light with the battery symbol 

flashes quickly and red  

(100 ms light/300 ms pause)

No acoustic signal

Detecting 

a child placed

 on the pad 

(evaluated in standby mode)

The indicator light with the power-on 

symbol flashes orange (1 Hz, 500 ms 

light/500 ms pause)

1x beep - warning after 30 secs 

from child presence detection, if the 

child is still detected (1 sec tone)

Detecting 

a child removed

 from the 

pad (evaluated in active mode)

The indicator light with the power-on 

symbol flashes orange

(5 Hz, 100 ms light/100 ms pause) (Ends 

when breathing detected, but 30 times 

at most.)

3 notification beeps (300 ms 

tone/300 ms pause)

(ends when breathing detected, but 

3 times at most)

Low temperature

 indication 

(evaluated in active mode)

The indicator light with the thermometer 

symbol flashes blue

(100 ms light/2 secs pause). 

(Lasts until the low temperature is no

longer detected.)

No acoustic signal

High temperature

 indication

(evaluated in active mode)

The indicator light with the thermometer 

symbol flashes orange

(100 ms light/2 secs pause). 

(Lasts until the high temperature is no

longer detected.)

No acoustic signal

Indication that the

 sensor pad

 is 

disconnected

 in standby mode 

(evaluated in standby mode)

3x red flashes in a circle around the 

 mechanical switch  

(100 ms light/100 ms pause)

1x beep - warning (500 ms tone)

Mechanical button jam

 detected in 

standby mode (evaluated in standby 

mode)

Light with the orange power-on symbol 

constantly lit. Releasing the button stops 

the light.

No acoustic signal

Error indication and warnings after 

automatic functionality test

 

(evaluated and indicated when swit-

ching to active mode)

Critical error detected - Red LEDs flash in a 

circle around the mechanical button

Warning detected - no light indication.

Acoustic indication according 

to the type of error or warning

Indication to 

enable/disable light

functionality

3x flashes of the lamp

(300 ms light/300 ms pause)

No acoustic signal

Indication of the 

Switch Guard  

function being 

disabled

3x orange flashes of the indicator 

light with the power-on symbol

(500 ms light/500 ms pause)

2x notification beeps 

(1 sec tone, 1 sec pause)

Indication of the 

Switch Guard

 

function being 

enabled

2x orange flashes of the indicator light 

with the power-on symbol

(1 sec light/1 sec pause).

1x notification beep (3 sec tone)

Indication of

 unsuccessful light 

activation

 or 

light forcibly turned off 

due to 

low/critically low batteries

5 x flashes of the indicator light with the 

battery symbol in red (100 ms light/100 

ms pause)

No acoustic signal

Summary of Contents for BM-03

Page 1: ...BM 03 Baby breathing monitor Certified medical device...

Page 2: ...aning 8 Alarm states 9 Operator warning 10 Important warning 11 Problems and how to solve them 12 Technical specifications 13 Definitions and symbols 14 EMC of a medical device 14 1 EMISSIONS limits b...

Page 3: ...ants for example due to sudden infant death syndrome SIDS or due to other causes suffocation an illness etc Due to the nature and purpose of use no side effects and other contraindications are known T...

Page 4: ...ONITOR Based on signals from the sensor pad located under the baby the device monitors the regularity of breathing and signals respiratory arrest Inhalation and exhalation are indicated by a green fla...

Page 5: ...bsence or presence Functionality test results A All indicators flash and short beep all checks were successful and the device is fully functional B Flashing lights and repeated warning beeps error dia...

Page 6: ...hich is mainly used to check on the baby at night Activation and deactivation of the light function The lamp function is deactivated by default To activate it simultaneously press the mechanical butto...

Page 7: ...t day and night or darkness Thanks to this the light diodes light up at night with a lower intensity than during the day to ensure parents or other caregivers have a good night s rest Day and Night mo...

Page 8: ...onnecting the connecting cable to the control unit Neither the sensor pad nor the control unit need to be specially cleaned or disinfected after unpacking Before using the monitor make sure that all p...

Page 9: ...dhesive tape to place it on the side of furniture or another solid surface cot clip a stand for the control unit it can be placed for example on the bedside table next to the bed Fig 10 Holder and Vel...

Page 10: ...eliminated in order to ensure the device works reliably and the baby is safe Disturbance can be caused by intense airflow fans air conditioning walking near the bed mechanical vibration from househol...

Page 11: ...nd the mechanical button will flash 3 times It is best to buy a set of 2 pads The second pad can be used in multiple places in the baby s first months of life for example in another bed at its grandpa...

Page 12: ...the mattress 180 in the cot once in a while or turn it upside down or let it air etc Only use a slightly damp cloth for cleaning Avoid wet wipes or other materials that may have loose fibres that can...

Page 13: ...n with the pad evaluated in active mode Moderate Orange colour flashes in a circle around the mechanical button 5 Hz 100 ms light 100 ms pause Technical alarm high intensity intermittent tone 100 ms t...

Page 14: ...use Lasts until the high temperature is no longer detected No acoustic signal Indication that the sensor pad is disconnected in standby mode evaluated in standby mode 3x red flashes in a circle around...

Page 15: ...nd must be placed at least 0 5 meters away from the child s head to prevent possible damage to its hearing Correct placement of the cot The device uses a very sensitive sensor to sense breathing Its o...

Page 16: ...eration due to the various types of equipment transmission powers and antenna systems Baby supervision help within reach Please note that the device can only alert you but it does not of itself preven...

Page 17: ...ance in the surroundings that need to be removed Follow section 4 After switching on the device indicates empty batteries Make sure you have not used rechargeable batteries they have a lower voltage a...

Page 18: ...tion IP31 Life span of the sensor pad 2 years from putting into service Service life of the control unit 10 years Service life of the device 10 years under the condition of replacing the sensor pad af...

Page 19: ...1 EMISSIONS LIMITS BY ENVIRONMENT Phenomenon Professional medical facilities a Home health care a RF emissions propagated by conduction and radiation CISPR 11 CISPR 11 c d Harmonic distortion see IEC...

Page 20: ...ces and ME systems with magnetically sensitive components or circuits e During the test the ME instrument or ME system may be supplied with any nominal input voltage but at the same frequency as the t...

Page 21: ...1990 GSM 1800 1900 CDMA 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulationb 217 Hz 2 0 3 28 1845 1970 2450 2400 to 2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Pulse modulationb 217 Hz 2 0 3 2...

Page 22: ...A new life begins MXA51302...