24
CAUTION
• Be sure to use sterilizers that can perform sterilization up
to 135°C. In some sterilizers, the chamber temperature
may exceed 135°C. Do not use these sterilizers as failure
of the handpiece could occur. Contact the sterilizer
manufacturer for detailed information about cycle
temperatures.
• Steam sterilization is recommended for the product. The
validity of other sterilization methods (such as plasma
sterilization or EOG sterilization) is not confirmed.
NOTICE
• The Wrench can be sterilized in a steam sterilizer at a
temperature not exceeding 135°C.
8-5
Storage
Store the product in a dry, clean location.
CAUTION
• After the sterization and drying cycles are complete,
remove the handpiece immediately from the sterilizer to
store it.
• Store the product in a well ventilated place out of direct
sunlight and within the range of temperature, humidity and
pressure specified in “11 Specifications”.
• Sterilization is not guaranteed after the sterilization
retention period specified by the manufacturer and seller of
the sterilization pouch has elapsed.
If the sterilization retention period has elapsed, perform
sterilization again with a new sterilization pouch.