FR
3) Perform main wash under the following conditions. Use a
Getinge 46 Series thermo-disinfector (or equivalent device):
Pre-Wash 1: Cold water (<110°F (43°C)) for 2 min
Pre-Wash 2: None - Not Applicable
Enzyme Wash: Hot tap water and 60ml of heated (95°F
(35°C)) detergent (Getinge Clean Renuzyme) for 3 min at
120°F (49°C)
Detergent Wash: None - Not Applicable
Rinse 1: Hot tap water for 2 min
Rinse 2: Hot tap water for 2 min
RO Final Rinse: Hot tap water for 32 min at 167°F (75°C)
Drying: 20 min
Upon removal from the thermo-disinfector, if the residual
moisture is present, wipe off with dry cloth or blow it off
with compressed air (≤0.35MPa) until there is no moisture
in the interior and exterior.
4) After cleaning and drying the handpiece, confirm that it is
clean under appropriate lighting (500 lx or higher).
If dirt still remains, repeat the procedures.
5) Proceed to “8-3 Cleaning and Lubrication”.
CAUTION
• Handpieces must be removed from the thermo-disinfector
immediately (within 1 hour) after the cleaning and drying
cycle is complete to prevent corrosion.
• After washing with thermo-disinfector, prior to lubrication,
dry the product until all internal moisture is completely
removed. Thermo-disinfector moisture remaining inside the
product could reduce the effect of lubrication and could
cause corrosion inside of the product.