![Nakanishi E3000C Operation Manual Download Page 16](http://html.mh-extra.com/html/nakanishi/e3000c/e3000c_operation-manual_704396016.webp)
29
30
Fig. 36
Connector A
14
15
16
17
18
19
20
21
22
23
24
25
1
2
3
4
5
6
7
8
9
10
11
12
13
4. Input/Output Connector A, B Pin Configuration
Connector B
9
10
11
12
13
14
15
1
2
3
4
5
6
7
8
1. Warning Function
Always check the control unit, motor, spindle and the condition of the cooling air prior to
use. This will help prevent system errors that will result in undesired operating conditions.
(1) The Warning LED [
20
] will flash.
(2) The Warning Code (listed in Table 2) will be displayed on the Digital Speed Indicator [11].
(3) A Warning Signal is output to the Warning Signal (PIN No. 20 : WARNING) of
Input/Output Connector A.
2. Detection of unsafe operating conditions
Always check the control unit, motor, spindle and the condition of the cooling air prior to
use. This will help prevent system errors that will result in undesired operating conditions.
When an Error Occurs the following events happen :
(1) Motor stops
(2) The Error LED [
19
] will flash.
(3) Error Code (listed in Table 3) will be displayed on the Digital Speed Indicator [
11
].
(4) An Error signal is output to the Error Signal (PIN No. 8 : ERR) of Input/Output Connector.
PROTECT FUNCTION
Warning Code
Warning Function
Trouble
A0
A1
A2
A3
Motor Cord
Low Air Pressure
Control Unit Overheat
Over Load
Internal Temperature of the Control Unit too High
Motor Torque load exceeding safe limits
A4
A5
Emergency Stop
Over Air Pressure
Emergency Stop system activated
Over Air Supply
Motor Cord or Connector is disconnected or misaligned
Low Air Pressure
Table 2
CAUTION
When the Warning LED
[
20
]
flashes, please check the Warning code.
Safety Relay Signal
PIN No. 3-11, 4-12
• The Safety Relay will be ON or OFF depending on the state of the Emergency Stop Signal
PIN1-9.
• When there is continuity between PIN3 (SAFE-1A) and PIN11 (SAFE-1B) or between PIN4
(SAFE-2A) and PIN12 (SAFE-2B) the motor is off. If there is no continuity between these
pairs of pins then the system is operating normally.
• If the Emergency Stop Signal is OFF (Open) the Safety Relay will be OFF (Open) and the
motor power will be interrupted and the motor will stop.
• If the ‘a’ contacts of the Safety Relay are welded together by an over load or short circuit
the ‘b’ contacts’ separation are maintained with more than 0.5mm spacing by the relay’s
recoil mechanism.
• The voltage/current specifications of PIN3-11 & PIN4-12.
Applied Voltage (V max) 30VDC
Working Current (Ip) 2A
Fig. 35
Input/Output Connector A
Plug Part Number : XM2A-2501 OMRON (or other similar high-quality product)
Cover Part Number : XM2S-2511 OMRON (or other similar high-quality product)
Input/Output Connector B
Plug Part Number : XM2A-1501 OMRON (or other similar high-quality product)
Cover Part Number : XM2S-1511 OMRON (or other similar high-quality product)
• The Plug and Cover are not provided with the system. Please purchase the specified plug
and cover from local suppliers.
• Use only shielded cables to minimize RF interference and noise. Connect the shield to
the plug cover.
• Different makers use different names for the cover.
CAUTION
To minimize RF interference and noise please keep the length of the cables as short as
practical and route separate from power cables.
3 (SAFE-1A)
4 (SAFE-2A)
12 (SAFE-2B)
11 (SAFE-1B)
The side of unit
Safety Relay
Contact
Motor
output part
Motor
output line
3. Input/Output Signal Connector Specifications
17
3
1
2
E3000C̲K0446E̲e̲080108 08.1.8 6:34 PM ページ29