background image

10

CAUTION

Use of dry air
Using compressed air containing excessive moisture could result in malfunction or failure 
of the motor and spindle. If excessive moisture or condensation is found in Air Filter Bowl, it 
will be necisary to install a larger Air Filter or dryer on the primary side of the Air Line Kit to 
prevent and remove excessive moisture.

Draining
Be sure to drain moisture and condensation from Air Filter Bowl regularly to avoid moisture 
being carried to Motor and Spindle.

Installation Location
Place this Air Line Kit on a flat and level surface.
If mounting on a wall, check if the wall is flat, and securely mount the Air Line Kit in a 
horizontal direction.

Handling
When removing the bowl for cleaning, carefully remove the bowl and do not damage to the 
bowl.

CAUTIONS FOR STORAGE, INSTALLATION and OPERATION

Do not use the Air Line Kit where corrosive gasses, chemicals, seawater, water, oils or steam 
exist.

Do not place in direct sunlight.

Do not use where the Air Line Kit is subject to vibration or repetitive shock.

Do not use where a heat source or radiated heat exist.

To conform to "STORAGE, INSTALLATION and OPERATION"  (Refer to " 3. SPECIFICATIONS ".)

This Air Line Kit is designed to supply clean and dry cooling and purging air to CONTROLLER and Motor / 
Spindle. 

The Air Filter traps small amounts of water and impure substances from the input air supply. This Air Line Kit is 
not intended to be used or replace an Air Dryer.

Air Pressure Regulator supplies steady airflow at a regulated air pressure.

The Air Line Kit can be wall mounted.

2

. FEATURES

Summary of Contents for AL-C1204

Page 1: ...essure exceeding 1 0MPa may cause the hose to rupture Inlet and Outlet Parts Do not hit impact or cause shock to the Inlet or Outlet Connector Parts Never put undo stress or load on the Inlet or Outlet Connector Parts Any damage to these components can cause air leakage and the inability of the outlet quick disconnect to adequately secure the tubing Dropping the Air Line Kit When installing the Ai...

Page 2: ...ling When removing the bowl for cleaning carefully remove the bowl and do not damage to the bowl CAUTIONS FOR STORAGE INSTALLATION and OPERATION Do not use the Air Line Kit where corrosive gasses chemicals seawater water oils or steam exist Do not place in direct sunlight Do not use where the Air Line Kit is subject to vibration or repetitive shock Do not use where a heat source or radiated heat e...

Page 3: ...ain Reservoir Capacity 80cm3 Safe Pressure Regulator Operating Range 0 05 0 85MPa Pressure Relief Automatic Relief Valve Dimensions W230 mm D120 mm H220 mm Weight 1 9Kg OPERATING ENVIRONMENT Temperature 5 40 C Humidity MAX 85 Atmospheric pressure 700 1 060 hPa TRANSPORTATION and STORAGE ENVIRONMENT Temperature 10 60 C Humidity 10 80 Atmospheric pressure 500 1060 hPa Standard Accessories Connection...

Page 4: ...eplacable by the end user Refer to Table 1 and Table 2 Table 1 Name Model manufacturer Air Filter M2000 8 W S CKD Regulator R2000 8 W Table 2 Air Filter Replacing Optional Name Model manufacturer Air Filter Bowl Assy Polycarbonate Bowl with Manual Cock F2000 W BOWL CKD Air Filter Consumable Part Kit The set of the O Ring Mantle and Bowl O Ring M2000 KIT S ...

Page 5: ...top draining To divert the drain to another location Connect a inside diameter ȭ6mm hose Not included provided by the end user to the Drain Output Port and divert to another location Lock and Release of the Regulator Knob Regulator Knob is equipped with Lock mechanism Lock Push the Regulator Knob IN to the Lock position Release Pull the Regulator Knob OUT to unlock Adjusting Air Pressure Turn the ...

Page 6: ... Fig 5 To Air Compressor Primary Joint Air Input Joint Motor EM 3030J EM 3060J E3000 CONTROLLER Secondary Joint Air Output Joint Mounting the Air Line Kit Bottom Mounting When the Air Line Kit is mounted from the bottom in a horizontal position Affix the 4 four Rubber Foot that are provided to the bottom Air Line Kit Rear Mounting Attach the Air Line Kit M6 Mounting Screws that are provided When Re...

Page 7: ...position Air leakage Check all threaded joints and re tighten if necissary Low air pressure Check the Compressor Air Circuit and Regulator Water dirt and debris are collected in the Air Filter Drain water dirt and debris from the Air Filter bowl 7 TROUBLESHOOTING If a problem or concern occur please check the following items prior to consulting your dealer 8 DISPOSAL OF THE Air Line Kit When dispo...

Page 8: ...2013 01 20 005 Ⓢ ...