User’s manual for LGB004 

 

Version 1, July 2023 

*Before receiving the instrument, please observe whether there are 
scratches, screws falling off, and other phenomena during 
transportation! 
*Do not press the test tube pad tightly for a long time before the machine 
is in an unused state! 
*It is strictly prohibited to install the instrument on uneven, shaking 
and vibrating workbenches! 
*It is strictly prohibited to perform shaking operations when the sample 
capacity exceeds 70% of the test tube capacity. To prevent liquid overflow, 
the container containing the sample must be covered! 
*The locking nuts on the left and right sides of the test tube fixing 
pressure plate must be tightened! 
*When installing the fixed pressure plate and sponge pad of the test tube, 
it is necessary to confirm that they are installed in place to prevent 
the test tube and sponge pad from flying out during operation! 
*If an emergency occurs during use, immediately press the emergency stop 
button! 
*After using the instrument, please cut off the power!                         
4.1【Introduction to the Installation of Test Tube Mats】              

            

Place the instrument in a suitable operating area, plug in the power, ensure that 

the power is in the OFF state, and select the required test tube pad as shown in 

the above image. 

The operation method is as follows:

】 

1. Firstly, loosen the locking nuts on both sides, remove the upper cover 

pressing plate, and then take out the test tube pad; 

2. Place the lower test tube pad flat in the test tube fixing plate; 

3. Select the desired test tube pad and place it flat and centered on the lower 

test tube pad; 

4. Select a test tube or container with a lid and place it evenly in the test tube 

pad; 

5. Place the upper test tube pad evenly above the test tube or container in a 

Summary of Contents for LGB004

Page 1: ...tube vortex mixer safety rules For safety reasons please read the operating manual carefully before starting the operation Components subjected to heat humidity and biological changes can lead to fun...

Page 2: ...ters 6 3 Introduction to the appearance and functions of the multi tube vortex mixer 7 8 4 Operating Instructions 8 11 4 1Introduction to the Installation of Test Tube Mats 9 4 2 Turn on off the power...

Page 3: ...ironmental impacts including natural ultraviolet radiation Corrosion and wear of protective cover parts and other safety parts Indoor use Altitude 2000m The applicable ambient temperature range of the...

Page 4: ...anual carefully before using the instrument Note Danger of high voltage Danger of electric shock Meaning of safety instructions In order to avoid harm to the surrounding people or damage to surroundin...

Page 5: ...face When installing the gasket or fixture make sure that it is in place to prevent the gasket and fixture from flying out during work Do not oscillate more than the maximum load of the instrument I P...

Page 6: ...s drug analysis Drug development and other scientific fields in food safety It shows incomparable advantages in handling a large number of sample mixing saving time and effort Automatic mixing improve...

Page 7: ...hout the need to adjust cumbersome parameters 6 The perfect optimization design of the body structure ensures sufficient counterweight weight and circumferential eccentric balance operation providing...

Page 8: ...lation mode circumference Circumferential diameter 3 6mm Time function display Digital timing 1s 999min Continuous Countdown Jog positive timing vein Speed regulation function display Impulse intermit...

Page 9: ...ube pressing plate B Lock nut C Upper test tube pad D Test tube rack E Lower test tube pad F Test tube fixed disk G Operation panel H Power switch I Power cord socket The above image shows the LGB004...

Page 10: ...irm key after adjusting the parameters press this key to confirm 6 Pulse mode set for intermittent operation 7 Stop Start key 8 Click the button to momentarily oscillate 9 Speed display area unit r mi...

Page 11: ...essary to confirm that they are installed in place to prevent the test tube and sponge pad from flying out during operation If an emergency occurs during use immediately press the emergency stop butto...

Page 12: ...3 When using any test tube pad in LGB004 evenly place the sample to be mixed and switch to continuous mode jog mode or pulse mode 4 Select continuous mode operation press the MENU key and the space on...

Page 13: ...esolved please turn out the emergency stop button before starting the instrument again 5 Maintenance matters Do not attempt to clean the instrument when the power cord is plugged in or the power switc...

Page 14: ...sponge or cotton cloth and then use distilled water to clean the detergent Do not sprinkle or spray the rotor with water as the liquid may be carried away and cause corrosion Allow to dry upside down...

Page 15: ...ontainers with a diameter of no more than 10mm with a total of 54 holes Test tube gasket Standard configurat ion LGB004 2 Max 2500rpm Used to place test tubes and small containers with a diameter of n...

Page 16: ...estarted after the fault is resolved Table 3 Fault alarm information Fault codes Code Definition Troubleshooting E2 overspeed Speed up too fast lower the gear up E3 Emergency stop switch System emerge...

Page 17: ...s the requirements of AC100 250V The instrument has high vibration and noise 1 Check if the instrument is placed on a horizontal desktop 2 Check if the test tube pressure plate is tightly pressed 3 Ch...

Page 18: ...of this user manual The implementation of installation debugging moving positions adding accessories and maintenance work is the responsibility of the production unit or sales company authorized by th...

Page 19: ...LGB004 3 1 3 Additional test tube pad According to the order contract 4 Certificate of Conformity 1 5 Quality Assurance Page 1 6 Operating Instructions 1 7 Power cord 1 When unpacking please check in...

Page 20: ...English 20 User s manual for LGB004 Version 1 July 2023 Maintenance records...

Reviews: