Visualization of errors
In case an error appears the centrifuge automatically stops. The error number will be
shown on display
(2.13)
(see Error chart).
Error chart
Security
The centrifuge must be used by previously qualified staff that knows how the equip-
ment works thanks to the user manual.
Put the centrifuge on top of a horizontal, plane and stable table making a free space at
least at 30 cm per side. The ventilation grilles must be always completely free of obstruc-
tions.
Do not put the centrifuge near any warm supply (burners, blowlamps…), nor expose it
directly to the sun, etc. Avoid vibrations, dust and dry environments.
During operation, dangerous material such as flammable liquids or pathological mate-
rial should be out of the working area.
Nahita centrifuges are not suitable for treatment of flammable liquids, explosive sam-
ples or samples capable of forming explosive mixtures.
The centrifuge is provided with a safety switch that avoids operation if the lid is open
and that in case of opening the lid during operation automatically stops centrifugation. It
is also provided with an electromagnetic lock that locks the lid when the rotor starts spin-
ning. Never try to open the lid while centrifuge is operating.
When you are not using the centrifuge for a long period of time please make sure it is
unplugged in order to avoid possible accidents.
It is essential to have the equipment switched off and unplugged from the net before
cleaning, checking components or replacing any piece (e.g. replacement of a fuse).
Never try to repair the centrifuge by yourself, since you will lose the warranty and may
provoke damages to the general operating system or the electrical installation, as well as
injuries to the people that usually handle the equipment (burns, hurts…).
Try not to spill any liquid on the control panel, though it is properly insulated. In case
you have any doubts do please immediately unplug the centrifuge (see Maintenance).
Made under the European regulations for electrical security, electromagnetic compati-
bility and security on machines.
Number
Type of error
E1
Unbalance
E2
Spin speed control loss
E3
Starting up with lid open
E4
Lid opening during operation
E5
Rotor missing or placed wrongly
E6
FCR or rpm too high for the rotor used
E7
Impossibility in detecting rotor (magnet loss)
Page 24
Instruction manual 52818000
Version 1 Jul.-08
Version 1 Jul-08
Instruction manual 52818000
Page 17
2.1 State indicator
2.2 Acceleration rate display
2.3 Acceleration rate selection button
2.4 Braking rate display
2.5 Braking rate selection button
2.6 Speed display
2.7 Speed adjustment buttons
2.8 FCR/rpm selection button
2.9 Time display
2.10 Time adjustment buttons
2.11 Temperature display
2.12 Temperature adjustment buttons
2.13 Rotor/Error display
2.14 Program selection buttons
2.15 Lid lock indicator
2.16 Pre-cooling function button
2.17 Quick spin function button
2.18 Start
2.19 Stop
Panel of controls
ENGLISH
ENGLISH
efrigerated Centrifuge
Model GL-18B
1.1
1.3
1.5
1.2
1.4
Run
Pre cool
Set out
1
2
3
4
5
RPM
RCF
xg
Pree-cool
Quick spin
Run
Stop
Program
X
R
E
min
ºC
1
3
4
1
2.18
2.5
2.14
2.17
2.16
2.7
2.11
2.13
2.9
2.3
2.4
2.2
2.1
2.10
2.6
2.8
2.19
2.12
2.15