NB-205QM/205QMC(Ver. 1)                                                                                                    13/17 

Platform Installation 

At  each  corner  of  base  platform,  there  are  four  thumb  screws  supporters  to  mount  universal  platform  or 
spring rack platform. One spring rack platform is basically mounted on the base platform. If other dedicated 
platform or tube rack platform is needed to use, just take spring rack out and fit the desired platform in four 
supporters, thumb screws. 

Do  not  operate  shaker  with  an  unbalanced  load.  Platforms  should  be  loaded  for  optimum 
stability and operation. Do not lift up platform when lifting up shaker to take out from chamber.   

 

Note: Sample Loading Capacity on Shaker 

RPM capacity with some flask platform is as following: 

Flask Platform 

Total Weight(sample volume) 

Reachable Max RPM 

250 * 9 

4.65(250ml full) 

300 

4.20(200ml) 

300 

500 * 5 

4.6(500ml full) 

300 

3.05(200ml) 

300 

1000 * 2 

* 3 
* 4 

3.85(1000ml full)   

300 

5.15(1000ml full) 

295 

6.5(1000ml full) 

260 

1000 * 2 

* 3 
* 4 

3.10(600ml) 

300 

4.60(600ml) 

285 

4.95(600ml) 

250 

2000 * 1 

3.4(2000ml full) 

300 

2000 * 1 

2.6(1400ml) 

300 

 
Do not operate the shaker at speeds that will cause the contents of vessels to be thrown out.   

 

 

Reversible Door Installation 

Door is installed to open basically from right to left. If it is needed to reverse the door open position, unscrew 
2  screws  of  door  hinge  that  are  fixed  in  main  body.  Leave  other  two  screws  in  hinge  which  fixing  door. 
Remove four plastic screws which are filled in the screw taps at right side(at front view) of incubator, fix the 
door hinge with securing screws in its screw taps.       

 

Drain hose Installation (NB-205QMC only)   

For NB-205QMC,  find the drain hose  at back of incubator.  Place it in proper place  avoiding area of  electric 
part. Place the tray at end of tube to collect water from chamber. Regularly check the collecting tray not to 
overflow. 

 
 

Summary of Contents for NB-205QM

Page 1: ...www n biotek com N BIOTEK NB 205QM QMC Ver 205QMQMC001 1 17 Stackable Incubator Shaker NB 205QM QMC Version 1 OPERATING MANUAL...

Page 2: ...1 3 Precaution for use 5 2 FEATURE SPECIFICATION 2 1 Features 6 2 2 Specification 6 3 CONFIGURATION CONTROL PANEL 3 1 Configuration 7 3 2 Stacking 8 3 3 Reversible Shaking Motion Platform Optional 9...

Page 3: ...the faulty caused by user s improper operation excessive use use of incorrect voltage frequency storage in wrong environment mentioned in manual 4 Complete the above table after installation and keep...

Page 4: ...essed by heavy products When it is damaged it can cause electric leakage or fire Do not handle the power cord with wet hands It can cause an electric shock Do not use the damaged power cord and outlet...

Page 5: ...incubator To prevent water and experiment material from going into the control panel during the experiment make sure to clean the control panel with a dry towel It can cause a fire and an explosion D...

Page 6: ...D 4W LED Operating Temperature 5 C 35 C 5 C 35 C Display Digital Display 5 Digit LED Digital Display 5 Digit LED Shaker One Mini Shaker Built in One Mini Shaker Built in Shaking Motion Orbital Orbital...

Page 7: ...Thermal Cut Off Power Cord Receptacle with Fuse Holder Glass Window Control Panel Screw Tap for Hinge of door for reversible door Shaker Photo 1 Front View Connector of Shaker Vent Holes To cool down...

Page 8: ...above cover you can find two kits with its screw tighten Install the stacking kit like Photo 4 Turn top cover over and fit it into the frame at top of incubator Photo 5 Locked with Stacking Kit Photo...

Page 9: ...e Shaking Motion Platform Optional Photo 6 Reciprocating Motion Type Photo 7 Orbital Motion Type 3 4 Shaker Connector Photo 8 Connector Socket of Shaker At Right and Back of Internal Chamber Photo 9 F...

Page 10: ...MP ON OFF Internal LED Lamp ON OFF Button MUTE Alarm Mute Button SET TEMP Temperature Set up Button Button for Increasing Value in Temperature Setting Mode Button for Decreasing Value in Temperature S...

Page 11: ...Lamp ON OFF Button MUTE Alarm Mute Button COOL Indication for Peltier Cooling SET TEMP Temperature Set up Button Button for Increasing Value in Temperature Setting Mode Button for Decreasing Value in...

Page 12: ...t base incubator is level and placed at bottom securely Open top cover of base incubator and find the stacking kits Referring some procedures and photos in page 8 fix the cover to support upper incuba...

Page 13: ...5 4 6 500ml full 300 3 05 200ml 300 1000 2 3 4 3 85 1000ml full 300 5 15 1000ml full 295 6 5 1000ml full 260 1000 2 3 4 3 10 600ml 300 4 60 600ml 285 4 95 600ml 250 2000 1 3 4 2000ml full 300 2000 1 2...

Page 14: ...u do not need timer and do need the control for shaking just by On Off switch leave the time set at t 00 00 4 To set the desired mixing speed when shaker stop Press RPM TIME key to switch RPM display...

Page 15: ...m will automatically stop If mute key is pressed during the shaker alarm 10 minutes of alarm delay will be given If there is door open during RPM alarm shaker will stop and alarm system will be inacti...

Page 16: ...through this delay program Over Temperature Cut Off visual warning only Thermal cut off have sensor attached on exterior surface of chamber seperately from main temperature sensor This protects over h...

Page 17: ...f chamber Then lift the shaker up vertically grabbing shaker body and take it out carefully from the chamber Clean the chamber with soft cloths and 70 alcohol After that dry the chamber completely bef...

Reviews: