TROUBLESHOOTING
6. TROUBLESHOOTING
6.1 CRANIOSTAT ADJUSTMENT
Should the craniostat not be centred with respect to the sagittal laser, perform a mechanical
adjustment:
Proceed as follows:
1.
Check that craniostat is centred
2.
Close the 2 side tabs
3.
Remove lower cover. There are no screws, just pull the cover downwards.
4.
Loosen the 2 screws shown in the figure and move the craniostat to align it to the sagittal laser.