6
Measurement
•
Make sure no dust gets into the instrument ports.
•
When not using the instrument for a long period of time, remove the limiting
aperture, and blow off dirt or dust on the protection glass with a blower
before use.
•
Over extended periods of use, changes in the surrounding environment may
result in the measured values shifting. Performing calibration periodically is
recommended to ensure measurements are accurate.
White Calibration Plate
•
Make sure the white calibration plate being used includes the same serial
number as that of the instrument being used.
•
The calibration data for the white calibration plate was measured at 23°C.
•
To achieve the highest accuracy when measuring absolute values
(colorimetric values), calibration and measurement should be performed at
23°C.
•
Do not allow the white calibration plate to become scratched or dirty.
•
When the calibration cap is not in use, keep the white calibration plate away
from external light and dust.
Ruler
•
The sliding surface has a special coating that facilitates sliding. If there is
dust or dirt on the ruler, use a blower to blow it off or gently wipe it with a
soft, clean dry cloth. Never use solvents such as thinner or naphtha.