Calibration
Calibration must be performed before measurement.
CAUTION
Calibration cannot be performed using just the instrument. Connect
the instrument to the PC beforehand, and start up the software.
Calibration
This instrument requires calibration before performing measurement after the
power is turned ON or if a certain amount of time has passed since the previous
calibration.
Calibration must also be performed after cleaning the white calibration plate or
the protection glass.
Memo
The reading may fluctuate slightly due to changes in the ambient
temperature or due to heat generation caused by repeated operation
of the instrument. In such cases, perform calibration regularly.
Calibration cap serial number
Serial numbers are included on both the instrument and the calibration cap.
Make sure the calibration cap being used includes the same serial number as
that of the instrument being used.
Memo
If the calibration cap was purchased as an optional accessory for
replacement, attach the included pairing number stickers to both the
instrument and the calibration cap, write the calibration data to the
instrument, and use the cap only for the instrument with the same
pairing number.
Memo
To write calibration data to the instrument, use the Spectrophotometer
Configuration Tool available for download from the following website.
https://www.myiro.com/downloads
Temperature conditions during calibration
Calibration must be performed at the same temperature at which measurement
will be performed.
Memo
Perform calibration after the instrument and the white calibration
plate have had enough time to adapt to the ambient temperature.