4. Dos and Dont’s
1. Use a grounding pin and make sure the power socket of the device is properly
connected before each operation.
2. Ensure that the voltage of the device is appropriate. If the voltage of the local
power supply is unstable, we recommend that users increase to the matching power
voltage.
3. In order to ensure the therapeutic effect and normal service life expectancy of the
device, please use the specified accessories provided or recommended by the original
manufacturer.
4. The device should not be placed in wet environment or near water, nor should
they be directly exposed to sunlight.
5. Do not place the device near a strong heat source, because this may affect the life
and normal use of the device.
6. Before treatment, please remove all metal objects from the body first, so as to
avoid unexpected conditions and influence the curative effect.
7. Please restrain from using the device on eyes, thyroid, parathyroid, testicles,
abdomen of pregnant women, and pacemakers.
8. Patients who are suffering from illness should consult with a doctor and gain
permission from the doctor before using.
9. Please turn off the power switch of the device, and ensure that the total power
supply of the device is turned off before the staff leaves after daily use, so as to
ensure the safety of the electrical products.
10. Use device or train device operators in strict accordance with instructions in the
user manual.
11. Do not aim directly at the knee when operating around the knee.
12. Make sure the probe is in full contact with the skin before operation.
13. Do not aim at the head when operating around the neck.
14. Do not use shock waves against hallow organs or the head.
15. The focus of the shockwave should not be used on spinal cord tissues, large blood
vessels and nerve pathways.
16. Shock waves should not be used on patients with coagulation disorders,
hemorrhagic diseases or untreated hemorrhagic diseases.
17. Avoid pressing the air outlet on the operating handle.
18. After the operation of shock wave, the machine must be shut down after the
exhaust fan on the handle stops cooling to avoid machine failure.
19. Gel must be applied on the operating area.
5. Troubleshooting & Solutions
1. The instrument cannot be started, and the key light on the back of the instrument
is not on?
A. Make sure the power cord is connected to a valid power outlet.