![MVE 21135284 Quick Reference Manual Download Page 14](http://html1.mh-extra.com/html/mve/21135284/21135284_quick-reference-manual_1860766014.webp)
MVE FUSION Quick Reference Guide
14 – 21205647 D
Calibration of Temperature Probe
The factory installed RTD temperature probe used on the MVE Fusion has been calibrated at the
factory using a two-point, low temperature range calibration method. This calibration method
provides a level of accuracy of +/- 1.8°F (+/- 1°C) when operating at altitudes between 1000ft to
1500ft (305m to 457m). Further calibration should not be required unless desired by the end
user. Refer to the MVE Fusion Technical Manual for information on calibration methods and
procedures.